9
5.16
Maximum /Minimum/ Average Value
Press
repeatedly
to
select
maximum,
minimum and average value of measure dta.
When MAX, MIN, and AVG are displayed at the
same time, it represents the elapsed time of the
measurement, and the present temperature T1 / T3.
5.17
Setup
Press
and hold for ≥ 2 seconds for SETUP:
Step1
~
Step4
Step1. High/Low limit Value setting for alarm
function (Limit):
Select Setup mode to the upper and lower limit of T1,
T2, T3 and T4 (Setup Limit Hi
、
Lo).
Press
to select and set the range (Hi & Lo) of
T1, T2, T3, T4.
Press
to select the position to be modified.
The position will be indicated to the next one by
each press.
Press
or
to set the plus-minus sign and
temperature.
The setting of the upper and the lower limits are
based on the TYPE range measured; it may
auto-identify whether it exceeds the limit. (The
Setup function will fail if it exceeds the TYPE range
which is set by the user.)
Press
to save the settings and proceed to
Step 2.