13
To test for continuity, connect the Meter as below:
1.
Insert the red test lead into the
HzV
terminal and
the black test lead into the
COM
terminal.
2.
Set the rotary switch to
and press
SELECT
button to select
measurement
mode.
3.
The buzzer does not sound if the circuit is
disconnected and resistance value is > 100 .
The buzzer sounds continuously if the circuit is in
good condition, with resistance value 10 .
4. The nearest circuit resistance value shows on
the display, the unit is .
Note
The LCD displays
OL
indicating the resistance of
the circuit being tested is higher than 400 .
The buzzer sounds if the
RESET
,
SELECT
or
is pressed.
When continuity testing has been completed,
disconnect the connection between the test
leads and the circuit under test.
Capacitance Measurement
(see figure 7)
The Meter’s capacitance ranges are: 40.00nF, 400.0nF,
4.000μF,
40.00μF,
and
100.0μF.
To
measure
capacitance, connect the Meter as follows:
1.
Insert the red test lead into the
HzV
terminal and
the black test lead into the
COM
terminal.
2. Set the rotary switch to
and press
SELECT
button to select
measurement
mode. The built-in equalized capacitance of the
Meter will affect the reading accuracy. For better
72-9375: OPERATING MANUAL