background image

7

Diode Test 

Perform the diode test to check diodes, transistors, and other semiconductor devices. 

The diode test sends a current through the semiconductor junction, and then measures 

the voltage drop across the junction. A good silicon junction drops between 0.5V and 

0.8V.

To test a diode out of a circuit, connect the meter as follows:

• 

Insert the red test lead into terminal and the black test lead into the COM terminal.

• 

Set the rotary switch to   to select diode test measurement mode.

• 

For forward voltage drop readings on any semiconductor component, place the 

red test lead on the component’s anode and place the black test lead on the 

component’s cathode. The measured value shows on the display.

• 

Note Connect the test leads to the proper terminals as said above to avoid error 

display. 

• 

The LCD will display OL indicating diode being tested is open or polarity is 

reversed. The unit of diode is Volt (V), displaying the forward voltage drop 

readings.

• 

Open circuit voltage is around 2.7V.

• 

When diode testing is completed, disconnect the test leads from the device under 

test.

Capacitance Measurement

To measure capacitance, connect the meter as follows:

• 

Use the multi-purpose socket and connect to the V and mA μA terminals.

• 

Set the rotary switch to   F.

• 

Connect the test leads across with the object being measured. The measured 

value shows on the display.

• 

When measuring capacitance values larger than 600uF, it is normal for the meter 

to require some time to stabilize.

• 

The LCD displays OL indicating the tested capacitor is shorted or it exceeds the 

maximum range.

• 

When capacitance measurement has been completed, disconnect the test leads 

from the device under test.

Note: 

The Meter displays a fixed value which is the value of the meters own internal 

circuitry. To ensure accuracy, it is necessary to subtract this value from the displayed 

value when measuring small capacitors.

Frequency Measurement

Warning: 

To avoid the risk of personal injury, do not attempt to measure frequency with 

voltage higher than 30V RMS.

To measure frequency, connect the meter as follows:

• 

Insert the red test lead into the Hz terminal and the black test lead into the COM 

terminal.

• 

Set the rotary switch to Hz to select Frequency measurement mode.

• 

Connect the test leads across with the object being measured. The measured 

value shows on the display.

• 

When frequency measurement is complete, disconnect the test leads from the 

circuit under test.

Summary of Contents for 72-8715

Page 1: ...1 Model 72 8715 Digital Bench Multimeter ...

Page 2: ... finger protection rings Ensure all inputs are less than the range selected otherwise it may cause electrical shock or meter damage Take caution when voltages are above 60V DC and 30V AC rms Do not use the meter with the battery or fuse covers removed Do not adjust the range selector during measurement Replace the batteries as soon as the low battery indicator appears on the display Remove dead ba...

Page 3: ...g altitude 2000m Maximum Display Digital 1999 Measurement Speed Updates 2 3 times second Polarity display Automatic Ω Function buttons Operation performed POWER Turn the power on and off LIGHT Turn display back light on and off HOLD To enter or exit range hold in any mode mV V Unit of Voltage The millivolt volt μA mA A Unit of current Microampere milliampere ampere Ω kΩ MΩ Unit of electrical resis...

Page 4: ...COM Diode Continuity Buzzer Measurement kHz V COM Frequency Measurement A mA μA COM mA μA DC Current Measurement 10A COM A DC Current Measurement A mA μA COM mA μA AC Current Measurement 10A COM A AC Current Measurement F V mA μA Use Multi Purpose Socket Capacitance Measurement ºC V mA μA Use Multi Purpose Socket Temperature Measurement hFE V mA μA Use Multi Purpose Socket Triode Enlargement Facto...

Page 5: ... equal to 10kΩ the error is negligible 0 1 or less DC or AC Current Measurement Warning Before connecting the meter in series with the circuit under test be sure power the circuit is off If the fuse burns out during measurement the meter may be damaged and there is risk of personal injury to the operator Use correct terminals function and range for the measurement When the test leads are connected...

Page 6: ... reading The LCD displays OL indicating open circuit or the resistor value is higher than the maximum range of the meter When resistance measurement has been completed disconnect the connection between the test leads and the circuit under test Continuity Test To test for continuity connect the meter as below Insert the red test lead into the Ω terminal and the black test lead into the COM terminal...

Page 7: ... capacitance connect the meter as follows Use the multi purpose socket and connect to the V and mA μA terminals Set the rotary switch to F Connect the test leads across with the object being measured The measured value shows on the display When measuring capacitance values larger than 600uF it is normal for the meter to require some time to stabilize The LCD displays OL indicating the tested capac...

Page 8: ...ment must between 18ºC to 28ºC to ensure accuracy especially when measuring low temperature Different reading may be obtained when testing room environment under short or open circuit situation then short circuited reading shall be considered as the correct reading Turning on the Display Backlight Warning In order to avoid the hazard arising from mistaken readings in low light situations please us...

Page 9: ...mV 0 8 3 750V Input impedance 10MΩ Frequency 45Hz to 400Hz Display True RMS 20V 10mV 200V 100mV 1000V 1V 1 0 4 DC Current 200μA 0 1μA 0 8 2 Fuse 500mA 250V fast type f5x20mm 2mA 1μA 20mA 10μA 200mA 0 1mA 10A 10mA 2 0 4 Fuse 10A 250V fast type f5x20mm AC Current 2mA 1μA 1 0 3 Fuse 500mA 250V fast type 5x20mm 20mA 10μA 200mA 0 1mA 10A 10mA 2 5 5 Fuse 10A 250V fast type 5x20mm ...

Page 10: ...o 20ºC 8 5 250V rms 20ºC to 0ºC 1 2 4 0ºC to 100ºC 1 2 4 100ºC to 1000ºC 2 5 10 Continuity Test Range Resolution Overload Protection Remarks 1Ω 250V rms Open circuit voltage approximate 3V When circuit disconnected with resistance value 100Ω buzzer does not sound When circuit is in good connection with resistance value 10Ω buzzer sounds continuously Diode Test 1mV 250V rms Open circuit voltage app...

Page 11: ...l type and specification as follows and make sure the fuse is inserted firmly in the holder Fuse 1 250mA 250V fast type 5x20mm AC220V Fuse 2 10A 250V fast type 5x20 mm A Fuse 3 250mA 250V auto recovery 5x20 mm μA mA Fuse 4 500mA 250V fast type 5x20 mm hFE Replacing the Battery Warning To avoid false readings replace the battery as soon as the battery indicator appears when using battery to power o...

Page 12: ...s is required Do not dispose of these items with general household waste Separate for the treatment recovery and recycling of the materials used Waste batteries can be returned to any waste battery recycling point which are provided by most battery retailers Contact your local authority for details of the battery and WEEE recycling schemes available in your area Made in China PR2 9PP Man Rev 1 0 ...

Reviews: