102
4.2.8
Comprehensive Example
First set the instrument to run in frequency sweep mode, and then set a square wave signal with amplitude of 1Vpp and
duty ratio of 50% as frequency sweep wave. The frequency sweep mode is linear. Set starting frequency to be 1kHz, stop
frequency to be 50kHz and frequency sweep time to be 2ms.
The specific steps are as follows:
Use frequency sweep function
Press SWEEP and Type in sequence to display a linear frequency sweep (press soft key Type to select) to start linear
frequency sweep function.
Figure 4- 126 Select SWEEP function
Select frequency sweep waveform
After linear frequency sweep function is used, press Square to select frequency sweep waveform, when the interface
below will display:
Figure 4- 127 Select frequency sweep waveform