Safe-T-Vue® 6 Instructions                                                                

 

Preparation

 

 Refrigerate Safe-T-Vue® for a minimum of 24 hours before activating. 

• 

Use a 

refrigerated cold pack 

beneath blood products to keep them  

      cold while dispensing (recommended). 

 

Handling

 

 Hold the Safe-T-Vue® indicator by the outer edges. 

• 

Do not hold or touch the temperature sensitive  

indicator areas, as your hands can pre-warm  
the indicator and affect its accuracy. 

 

Application

 

1. 

Apply the Safe-T-Vue® indicator to the blood bag: 

• When

 the blood is requisitioned and dispensed  

from the blood bank, OR 

• 

When the blood is received into your blood 

bank inventory 

Color Change Temperature Indication

 

• 

The white dot will begin to gradually fill in  

with red dots at approximately 5°C. 

• The

 appearance of small red dots (typically  

starting around 5°C) is an indication that the  
blood should be cooled or re-refrigerated  
immediately, before 6°C has been reached. 

• The

 temperature-sensitive dot on                      

Safe-T-Vue® 6 will turn from white to  
completely red when the blood product  
temperature reaches 6°C. 

• Extensive

 testing has shown that the color  

change from white to completely red should  
be clear and unambiguous if Safe-T-Vue® has  
been handled properly. 

 

Validation

 

 
 
 

 

< 5° C 

 
 
 
 

 

≈ 

5.5° C 

 
 
 
 

 

≥ 

6° C

2.

 

Remove the adhesive backing and apply  
Safe-T-Vue® to the area of the blood bag with the  

largest blood volume. 

3.

 

Be careful not to touch the center indicator areas. 

4

. (Optional)

 Some users may achieve more  

consistent results and prevent false positives by  
returning the blood bag to the refrigerator with  

Safe-T-Vue® attached for up to 15 minutes BEFORE  

activating. 

 

Activation

 

1.

 

It is preferable to activate Safe-T-Vue® in the  
blood refrigerator. 

2.

 

Pull back the foil to expose the indicator’s active 

 

areas. Again, be careful not to touch the center  
indicator areas. 

3. 

Activate the indicator by folding the rounds  

together and snap shut.

  

4. 

The blood is ready for transport to destination,  

or can be placed back into the refrigerator until  
needed. 

 