7. Maintenance & Trouble shooting

7.1 Maintenance

The calibration instrument has been designed with the utmost care. Ease of operation and simplicity

of maintenance have been a central theme in the product development. Therefore, with proper care

the instrument should require very little maintenance. Avoid operating the instrument in an oily,

wet, dirty, or dusty environment.

If the outside of the instrument becomes soiled, it may be wiped clean with a damp cloth and mild 

detergent. Do not use harsh chemicals on the surface which may damage the paint.

Avoid knocking or dropping the calibrator.

If the mains supply cord becomes damaged, replace it with a cord with the appropriate gauge wire 

for the current of the instrument. 

Depending on the environment in which it is used, periodic cleaning is recommended. Cleaning may

be accomplished by the use of a small dry paint brush.

 

                                                                

7.2 Trouble shooting
1. Unit fails to operate

Check fuse  if it is tripped switch is ON. If not power ON of calsys 1500 BB consult us.

 2.

 

Unit unstable

 Controller parameter has been interfered, consult us. 

3. If the temperature of the calibrator is not rising 

(a)

The heating element may be open.  

(b)

The thermocouple may be open. 

(c)

The SSR may be damaged.

(d)

The controller may not be giving output

(e)

The ambient temperature inside the chamber is raised and safety controller switched
OFF the power

 

CAUTIONARY NOTE   

TEMPSENS PRODUCTS ARE INTENDED FOR USE BY TECHNICALLY TRAINED AND COMPETENT
PERSONNEL FAMILIAR WITH GOOD MEASUREMENT PRACTICES.

IT IS EXPECTED THAT PERSONNEL USING THIS EQUIPMENT WILL BE COMPETENT WITH THE
MANAGEMENT   OF   APPARATUS   WHICH   MAY   BE   POWERED   OR   UNDER   EXTREMES   OF
TEMPERATURE, AND ARE ABLE TO APPRECIATE THE HAZARDS WHICH MAY BE ASSOCIATED
WITH, AND THE PRECAUTIONS TO BE TAKEN WITH, SUCH EQUIPMENT.

Summary of Contents for CALsys 1500 BB

Page 1: ...CALsys 1500 BB CALsys 1500 BB Black Body Temperature Calibrator USER MANUAL...

Page 2: ...on 3 2 Circuit diagram 4 Overview 4 4 1 Introduction 4 2 Outline description 5 Operating Instruction 5 5 1 Power 5 2 Setting the temperature 5 3 Operating Instruction 6 Operation of Controller 6 6 1 T...

Page 3: ...through defective material or workmanship provided the failure is not the result of misuse In the event of failure covered by this warranty the instrument must be returned carriage paid to the supplie...

Page 4: ...the flexible mains cable which is colored green yellow MUST be connected to a suitable earth Warning Internal mains voltage hazard Do not remove the panels There are no user serviceable parts inside C...

Page 5: ...230 V AC Power Maximum Supply Frequency 50 60 Hz Temperature range 500 to 1500 C Resolution 1 0 C Stability 1 0 C Controlling sensor R type T C Time to reach max Temperature 2 5 hours Operating Temper...

Page 6: ...other pro ducts DANGER Power is present even if the front panel circuit breaker is in the OFF 0 position 4 2 Outline Description Calsys 1500BB is a transportable unit designed for use on any reasonab...

Page 7: ...ed same sr no On heater holder and chamber NOTE HEATERS INSERT VERY CAREFULLY After Inserting its tight by screwdriver Open back panel sheet of CALsys 1500 and remove all Packing sheet from inside Ins...

Page 8: ...ven picture Open holding ceramic clamp as shown in figure Open the front plate of CALsys 1500BB and remove all packing sheets from inside Fix the 6 heater into the chamber and Connect all the aluminum...

Page 9: ...ensure that the tail of cavity fix to the groove on the back panel And make sure cavity properly insert at back panel hole as shown in image After insert the cavity tight the holding cap of SS chambe...

Page 10: ...5 1 Power Plug the black body power cord into a mains outlet of the proper voltage frequency and current capability Typically this will be 230 VAC 10 50 60 Hz Turn on the black body using Mains switc...

Page 11: ...n the span Furnace is stable when PV is equal to SV Set Value 6 Master sensor take some times to reach the setpoint temperature and stable at temperature near about controller set temperature 7 When t...

Page 12: ...the set point to the required value The lower display changes to indicate the new set point 6 3 Monitoring the Controller Status A row indicate the controllers status as follows OP1 Heat Output REM Th...

Page 13: ...s recommended Cleaning may be accomplished by the use of a small dry paint brush 7 2 Trouble shooting 1 Unit fails to operate Check fuse if it is tripped switch is ON If not power ON of calsys 1500 BB...

Page 14: ...f software 1 2 Parameter in mainscreen 1 2 1 Communication Communication between the furnace and the software is implemented via a RS 232 cable connected between the furnace and the PC serial port Thi...

Page 15: ...ging up to 120 minutes than clickon start graph button After complete the task clickon save to file button Set Point C in which you can set temperature of furnace as your requirement PVI Value Read th...

Page 16: ...el to protect it from humidity Warranty TEMPSENS CALsys 1500BB instrument have a warranty of one year from the invoice date This warranty covers manufacturing defects User induced faults are not cover...

Page 17: ......

Reviews: