TempraMed VIVI CAP 1 Quick Start Manual Download Page 1

Intended Use:

VIVI CAP is a reusable replacement cap for insulin pens. It is intended to keep 

insulin cool within safe temperatures below 30°C (86°F) for minimum of 12 

hours even in a constant environmental temperature of 37.8°C (100°F), when the 

device is placed inside a personal handbag. The cap temperature regulation is 

passive and combines thermal insulation with heat absorbent material. The cap is 

‘reactivated’/’filled up’ to provide safe storage conditions, whenever the ambient 

temperature gets below 26°C (78.8°F), for subsequent periods of 12 hours all 

year long.

Note that insulin must always be at the manufacturer’s recommended 

temperature. Please refer to the insulin package insert of your specific insulin pen. 

Device Components:

The device consists of the following components:

•  Thermal insulation (stainless steel), held between the outer plastic cover 

and the inner plastic adaptor. 

•  Inner  filling  of  Phase  Change  Material  (non-toxic  &  non-flammable  – 

Safety Data Sheet available if required), held between the thermal 

insulation and the inner plastic adaptor.

•  Inner plastic adaptor that fits the pen type

•  Outer plastic cover 

•  Primary (non-rechargeable) battery, switch, Green/Red indicator light and 

temperature sensor  

•   Ring shaped plastic pen-adaptors with breakable key code letters 

Operating Instructions:

1.  Release the cap from the package, press & release the switch [there is 

no need to continue pressing the switch] and check the indicator light. 

If the indicator light does not turn on, contact the manufacturer. DO NOT 

USE the device.

2.  If the indicator light is RED, DO NOT USE the device, leave it in an ambient 

temperature of less than 26°C (78.8°F) for at least 5 hours or place it 

inside a refrigerator for about an hour. Then, press & release the switch 

and check that the indicator light is GREEN.

3.  If after step 2 the indicator light is not GREEN, contact the manufacturer. 

DO NOT USE the device. 

4.  When the indicator light is GREEN, you are ready to use the activated VIVI 

CAP with your medication pen. But first you should fit the proper pen type 

ring adaptor to VIVI CAP. This procedure should be performed once only. 

5.  Choose the proper pen adaptor according to your pen name and the 

letters marked next to the plastic ring adaptor : F/S – for FlexPen® or 

SoloStar®, K-for KwikPen®, T-for FlexTouch®*. Remove your pen cap 

and slide the pen adaptor over your pen starting with the needle side 

while the adaptor letter faces the needle side (see Fig 1-3) until it locks 

with  a  “click”  sound”.  For  FlexTouch®*  note  that  the  Pen  is  aligned 

properly to the adaptor- Fig 2. If the adaptor doesn’t lock, make sure you 

have matched the pen adaptor to your pen. While holding the pen around 

the pen adaptor with fingers of one hand, with your other hand tilt “back 

and forth” the adaptor letter until it breaks off Fig 4. While aligning the 

grove on the adaptor to the rail on the top part of VIVI CAP Fig 5, insert the 

pen with the ring adaptor to VIVI CAP until it locks with a “click “sound 

Fig 6. The ring adaptor will connect “flat” to the top part of VIVI CAP, will 

remain inside VIVI CAP and will fit it to your  your pen

6.  Use your pen as prescribed.

7.  Always remove the used needle after each injection and store the insulin 

pen without a needle attached.

8.  During daily use, the indicator light should be checked by pressing the 

switch and observing the indicator light color. Even if the indicator light 

is  GREEN,  to  assure  proper  operation,  the  device  should  be  placed  for 

about 6 hours in an ambient temperature of less 26°C (78.8°F) at least 

once every 12 hours.

9.  At any time if the indicator light is RED when checking it, stop using the 

device. Place the device in an ambient temperature of less than 26°C 

(78.8°F)  and  start  using  it  again  when  while  checking  it,  the  GREEN 

indicator light has turned on.

10. If  during  normal  use,  after  1  year  from  first  activation  while  in  room 

temperature, the indicator light does not turn on, the battery is depleted. 

The cap must be disposed of properly.

Device Signals:

The device has one indicator light, which changes color according to the 

system status, after the switch is pressed and released:

Maintenance Troubleshooting & Disposal

WARNING:  The  device  contains  no  user-serviceable  parts.  Do  NOT  try  to 

open the cap. In any case of malfunction contact TempraMed.

You  can  clean  the  exterior  and  interior  of  VIVI  CAP  by  wiping  it  with  a 

medicinal swab. Do not soak, wash or lubricate VIVI CAP, as it may leave 

residue on the interior of the cap

Dispose of the device according to local regulations for electrical/electronic 

waste

 

Storage and handling

•  The VIVI CAP is designed to work accurately and safely. It must be handled 

with care. If dropped or crushed, there is a risk of damage to the insulation, 

leakage of phase change material or damage to the indicator light. 

•  Do not use the VIVI CAP with insulin pen if there is no indicator light, while 

in room temperature, or if the indicator light shows RED or if the cap is 

broken or otherwise damaged.

•  Do not store the VIVI CAP exposed to water or other liquids. 

•  VIVI CAP should only be stored at temperatures between  -10°C to 60°C

Device Physical Specifications:

Length 90mm (3.5”), Diameter 30mm (1.18”), Weight <70gr (2.5oz)Contains 

1x3V primary battery

Support and Contact Information

For answers to questions regarding the use of the VIVI CAP contact:

TempraMed Israel Ltd. www.my-vivi.com

AMAR registration No 30240001 by TempraMed Israel Ltd. 64 Heh B’Iyar Tel Aviv 

62198. Tel: 03 9195335

E mail - [email protected]  Website - www.my-vivi.com

VIVI CAP is Patent Pending: PCT/US2015/040655 and other patent 

applications.

VIVI CAP, the VIVI logo, TempraMed, TempraMed logo, are TM ® and ©, 

TempraMed Ltd. 2016.

Warranty

: This product is guaranteed against manufacturing defects for 

a warranty period of one year from the date of purchase. In the event of 

any mechanical problems, the product should be returned to the place of 

purchase with the relevant proof of purchase for a replacement within the 

warranty period

Disclaimer

 : These instructions are only a guide. Due to varying 

environmental conditions around the world, VIVI CAP will respond differently 

according to the location. 

Always follow the insulin manufacturer’s recommendations, regarding 

storage of insulin pen. The user maintains full responsibility regarding proper 

use and storage of insulin 

pen

.  TempraMed assumes no responsibility for 

incidental or consequential damages including and without limitation to 

damages or injuries resulting from product malfunction.

TempraMed’s liability will never exceed the purchase price paid by the user. 

Proof of purchase is required.

Document updated January 1, 2020

For further support please visit http://my-vivi.com

Copyright © TempraMed Inc., 2020. All rights reserved.

Document Number  2020

No part of this publication may be reproduced in any material form (including 

photocopying, or storing it in any medium by electronic means whether or 

not transiently or incidentally to some other use of this publication) without 

the prior written permission of the copyright owner, or under the terms of a 

license issued by the copyright owner.

The information contained in this document is subject to change without 

notice.  TempraMed  is  neither  responsible  for  nor  liable  to  anyone  in 

connection with this document.

*Registered trademarks of: Novo Nordisk, Sanofi, Eli Lilly

GREEN

RED 

No lights

Indicates proper operation and temperature below limit. 

Safe to use the device
Temperature above limit. 

Do not use the cap on the pen. Remove the pen from the cap. 

Place the cap in ambient temperature of less than 26°C (78.8°F) 

until while checking it, GREEN indicator turns on.
Battery is depleted, stop using the device and dispose properly.

VIVI CAP1

Insulin Temperature Shield

eng

Problem
The insulin pen does 

not fit to the cap 
No indicator lights 

RED indicator light 

Solution
Identify and use the correct pen ring adaptor  to match 

the pen model. 
1. At first use, do not use the product. 

Contact the manufacturer.

2. After and during use, the battery is depleted. 

Do not use the product and dispose of properly.
Stop using with the pen.

Place in refrigerator for about one hour or in an ambient 

temperature of less than 26°C (78.8°F) for at least 5 hours. 

Press and release the switch and check for a GREEN light.

If light is still RED, contact the manufacturer for support.  

VIVI CAP1

Insulin Temperature Shield

Switch

Indicator

Cooling to recommended temperature | Just put it on and it works

No need for refrigeration to activate | Temperature assurance indicator

Replacement to pen cap | Small and lightweight

Fits your pocket | One year

 

of daily use

Keeping insulin cool 365-day maintenance free Confidence on the go

Learn more at   www.my-vivi.com

Pen  Adaptors with code letters

Fig 1.

Push!

Push!

Fig 4.

Fig 2.

Attention!

For FlexTouch only

“Click” sound

“Click” sound

OK

Fig 5.

Fig 3.

Fig 6.

T

K

F S