Version 01 /
p.15
AFA1000/1 MK2 Airflow Monitor
Installation & Operating Manual
b. Coms data -
the Screen will show the coms setting data for the relevant selected
protocol:-
Protocol = None/TEL/Modbus/BACnet
ID = Slave ID for Modbus or Device Instance for BACnet
Baud Rate = Shows selected Baud Rate
Parity = Shows selected Parity
Tx & Rx = the display will show the current data packets sent and received, the displayed
value will rollover to zero when the maximum count is reached.
c. I/O Status -
when selected the following options are shown:-
Input Data
Output Data
Sensor Data
Done
Input Data:-
Input 1 - 0 / Off / On / Not Used
Input 2 - 0 / Off / On / Not Used
Input 3 - 0 / Off / On / Not Used
0 = Analogue Input e.g. Sash Position sensor Input Voltage (0-5VDC)
Off = Input Open
On = Input Closed
Not Used = Input not assigned
Output Data:-
Output 1 - Off / On
Output 2 - Off / On
Output 3 - Off / On
Output Off = Output Open or not assigned
Output On = Output Closed
Sensor Data:-
Airflow 00.0 %
Temperature
°
C
Airflow % = Output of airflow sensor in %, 100% = no airflow
Temperature = Temperature in
°
C or
°
F - (when enabled)
Done - returns to Diagnostics menu.
d. Done -
when selected the controller returns to the run screen.
Summary of Contents for AFA1000/1 MK2
Page 4: ...1 SAFETY...
Page 9: ...2 OVERVIEW OF THE AIRFLOW MONITOR...
Page 19: ...3 FUNCTIONS OPERATION...
Page 24: ...4 COMPONENTS...
Page 27: ...5 INSTALLATION...
Page 32: ...6 CONFIGURATION CALIBRATION...
Page 36: ...7 AUXILIARY FEATURES CONNECTIONS...
Page 41: ...8 RS485 COMS OUTPUT...
Page 48: ...9 WARRANTY...