Temco MAGIC II Installation And Calibration Download Page 2

MAGIC II 

Training Guide – 11/14 

 

Page 2  

SETTING UPPER & LOWER RANGE VALUE WITH HART COMMUNICATOR 

 

Connect one lead of the communicator to the positive terminal and one lead to the negative 
terminal, or connect the leads across the dropping resistor. 
 

1)

 

Connect with your HART communicator and then go to the LRV and URV screen. 

2)

 

Set the Lower Range Value (LRV) to match control room setting. 

3)

 

Set the Upper Range Value (URV) to match control room setting. 

 
 

FIRST TIME ZERO ADJUSTMENT  

1.

 

Put Consistency Loop in Automatic Control. 

2.

 

Obtain a Lab Sample.  Record the 

MAGIC’s

 signal at the time the sample is taken. 

3.

 

Adjust Coil so the 

MAGIC’s

 signal agrees with Lab Sample 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

SN Tag w/ 

Coil Voltage 

Voltage Test 

Points 

mA Meter 

Connection 

24 VDC Terminal  

Connection 

Summary of Contents for MAGIC II

Page 1: ...leg 5 Connect digital voltmeter to front panel test points With motor off adjust Coil to DC voltage marked on S N tag Magic box 6 Use HART Communicator to set the Upper Range Value URV and Lower Range Value LRV to match control rooms settings 7 TEMCO has preset the curve for your consistency application 8 Electronic dampening is factory preset Do Not Change 9 Switch HART to read Process Variable 1...

Page 2: ...mmunicator and then go to the LRV and URV screen 2 Set the Lower Range Value LRV to match control room setting 3 Set the Upper Range Value URV to match control room setting FIRST TIME ZERO ADJUSTMENT 1 Put Consistency Loop in Automatic Control 2 Obtain a Lab Sample Record the MAGIC s signal at the time the sample is taken 3 Adjust Coil so the MAGIC s signal agrees with Lab Sample SN Tag w Coil Vol...

Page 3: ...MAGIC II Training Guide 11 14 Page 3 MAGIC CURVE SELECTION Furnish MAGIC II Curves Unbleached Stock 10 OCC and Bleached 7 TMP and Newsprint 6 The curve is factory preset for your application ...

Page 4: ... refer to Ouput Adjustments to change zero Varying error with changes in consistency Refer to Output Adjustments to change sensitivity with the Curve Number MAGIC output does not change Voltage can vary 05 VDC Verify that breather plug on damper has been removed Verify that shipping brace has been removed Examine area around torque arm and motor for objects obstructing free movement Disconnect loo...

Page 5: ...late 20mA output ____VDC 20mA 5 Check Coil Voltage with MOTOR OFF ____VDC 6 Does Coil Voltage go UP when torque arm is pushed to the right and DOWN when pushed to the left ____ Yes ____ No 7 Does Consistency Output follow the coil voltage changes ____ Yes ____ No 8 Disconnect ThermaCoil from electronics enclosure Does Coil Voltage equal approximately 10vdc ____ Yes ____ No TEMCO Engineering Inc 16...

Reviews: