           DE-OX® SUN Helium  

analyzer Draft Issue n. 1.1 of 10/02/11    

  page

5  

mixed gas flow, gives an electrical signal. This signal is transmitted to the central unit and processed in order to give 
the  helium  percentage  information  on  the  display.  The  mixed  gas  flow  that  comes  in  contact  with  the  sensor,  must 
range from 0,5 to 1 litres/min (best 0,5 lit/min) at the same calibration pressure (typically ambient pressure 1 bar) and 
temperature (Typically 20°C) or a wrong gas analysis will occur (preferably use a flow regulator with meter connected 
to  the  mixed  gas  tank).  Higher  flow  value  or  pressure  or  temperature  variations,  will  cause  wrong  analysis  and  will 
irreparably  damage  the  sensor.  During  the  gas  analysis,  the  unit  must  be  kept  steady  and  must  not  be  shaken  or 
submitted to vibrations.  Allow gas freely flows into the pipes. 

 

During gas analysis, the helium percentage of the mixed gas being analyzed is available on the display. It is possible 
to make several gas analysis without waiting the instrument goes to 0 reading. 

 

It is possible to let 

DE-OX® SUN 

in ambient for ambient analysis of helium. In this case the reading will be slower (up 

to 5 minutes).  

Warning 

 

When  battery  is  nearly  exhausted  BAT  appears  on  the  display.  In  this  case  helium  analysis  is  not  anymore 
reliable. 

Warning 

 

Breathing  mixed  gas  with  helium  needs  the  right  training.  NOOA  and  medical  studies  recommend  not  to 
expose to an oxygen partial pressure higher than 1.6 bar or lower than 0.18 bar because of serious personal 
injury including death. Refer to proper medical literature and receive the right training. 

 

 

 

Replacement of the Battery and of the Helium Sensor 

 

DE-OX® SUN 

requires one 9 Volt transistor alkaline battery. It must be replaced when “bat” appears on the display. 

To replace the battery: 
-  Switch off 

DE-OX® SUN

.  

-  Remove  the  three  closing  screws  and  carefully  open  the 

instrument.  

-  Remove the old battery and replace it with the new one. 
-  Do  not  throw  anyway  the  old  battery.  Check  the  correct  recycling 

procedure. 

-  Close the instrument screwing the three closing screws. 
-  Switch on the instrument with the On/Off  key and check the battery 

level. 

Warning 

 

Don’t leave battery inside the instrument if you do not use it for 
a long time. 

 

 

 

The  helium  sensor  must  be  replaced  when  unable  to  correctly  read 
helium  concentration.  The  sensor  life  is  over  24  months  or  more 
under  normal  operating  conditions.  Always  contact 

TEMC

® 

for 

replacing the sensor. 

Warning 

 

Replace  the  helium  sensor  when  the  unit  is  unable  to  correctly  analyze  the    gas.  The  use  of  an  exhausted 
sensors may cause dangerous analysis and lead to death. Wrong helium analysis may lead to death. Always 
contact 

TEMC

® 

for sensor replacement. Do not use any other helium sensor brand. 

 

 

 

 

To replace the helium sensor: 

 

-  Switch off 

DE-OX® SUN 

with the On/Off  key. 

-  Remove the three closing screws and carefully open the instrument.  
-  Unplug the sensor connector. 
-  Remove the old sensor screwing un-clockwise the flow adapter and replace it with the new one. 
-  Connect again the plug to the sensor connector. Screw again the flow adapter. 

Summary of Contents for DE-OX SUN

Page 1: ...s 2 Powering On and Warming Up 3 Reading Page 3 Programming Procedures 3 Threshold Alarms Setting 3 Full Scale Value Setting 4 Auto Switch Off Delay Time 4 Threshold Alarms 4 Powering Off 4 Factory Reset 4 Note For Correct Helium Analysis 4 Replacement of the Battery and of the Helium Sensor 5 Central Unit Specifications 6 Helium Sensor Specifications 6 4 20mA Connections if available 7 Open Colle...

Page 2: ...ion The operator of this instrument is responsible for any damage or injury resulting from improper use unauthorized repair improper maintenance or damage by anyone other than TEMC Use always another instrument for data cross checking The N O A A National Oceanic and Atmospheric Administration oxygen partial pressure and exposure time limits table are available on world wide literature TEMC whose ...

Page 3: ... and the sensors must be firmly kept and must not be shaken The gas pipe used to provide gas mix to the sensor must allow free gas flowing Never seal or stop the sensor holes If this event occurs the helium sensor will be irreparably damaged Warning The mixed gas flow must be 0 5 2 litres min at ambient pressure typically 1 bar and temperature typically 20 C or DE OX SUN will yield a wrong gas ana...

Page 4: ...t will show the helium concentration Warning The instrument will switch off after the time set into the Auto Switch Off Delay Time If you are using the instrument in continuous blending system set this value to 000 If not the instrument will auto switch off without giving any control of the mixing process Always analyze the mixed gas you are going to dive with Diving with wrong mixed gas may lead ...

Page 5: ...gher than 1 6 bar or lower than 0 18 bar because of serious personal injury including death Refer to proper medical literature and receive the right training Replacement of the Battery and of the Helium Sensor DE OX SUN requires one 9 Volt transistor alkaline battery It must be replaced when bat appears on the display To replace the battery Switch off DE OX SUN Remove the three closing screws and ...

Page 6: ...e the instrument The instrument or any of its part are waterproof If one of these events occur the DE OX SUN will be irreparably damaged Helium Sensor Specifications The Helium sensor is developed for related underwater applications It operates by comparing the thermal conductivity of the sample gas with that of a reference gas in this case air A heated sensor is mounted so that it is exposed to t...

Page 7: ...Always contact TEMC for sensor replacement Do not use any other helium sensor brand 4 20mA Connections if available In one side of the instrument if equipped with this feature there is a jack for the 4 20 mA output Insert the plug into the jack for taking out the signal For connections refer to the following diagram Warn ing The plug should be connected or disconnected when the instrument is switc...

Page 8: ...ry Warranty will be void by failure to install use or maintain DE OX SUN according to TEMC instructions To avail oneself of the warranty send the product with carriage prepaid to TEMC THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES EXPRESSED IMPLIED OR LEGAL TEMC MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE Buyer s exclusive remedy shall be the repl...

Page 9: ...________n_____ Address Città______________________________CAP__________ Town ZIP Nazione_________________________________________ Country Telefono_________________________________________ Telephon E Mail_________________________________________ E Mail N serie strumento_________________________________ Instrument Serial number Data Acquisto____________________________________ Purchase date Dichiaro...

Page 10: ...er Draft Issue n 1 1 of 10 02 11 page10 Disposal of Old Electrical and Electronic Equipment In the EU when you buy a new equipment you may give back the old equivalent unit to the supplier free of charge for correct disposal ...

Page 11: ...il temc iol it Website www temc it DE OX SUN ANALIZZATORE DI ELIO Guida veloce di riferimento Attenzione Fare sempre riferimento al manuale completo in formato elettronico prima di utilizzare lo strumento Questa guida è un e veloce supporto all uso ma non è completa L uso non corretto di questo strumento può provocare la morte Per accendere lo strumento tenere premuto il tasto ON OFF per qualche s...

Page 12: ...DE OX SUN Helium analyzer Draft Issue n 1 1 of 10 02 11 page12 E Mail temc iol it Website www temc it ...

Reviews: