![TELEVAC MX7B Instruction Manual Download Page 18](http://html.mh-extra.com/html/televac/mx7b/mx7b_instruction-manual_1082545018.webp)
©2022 Televac® - The Fredericks Company | www.televac.com | [email protected] | +1 215 947 2500
mx7b_im rev E Page
18
of
23
9.4
RS-485 and USB Command List
9.4.1
RS-485 and USB Status Commands
Command
Response
Description
S1
ppsee
Read pressure data where ppsee is:
pp = mantissa
s = sign
ee = exponent
S2
yymmdd
Read firmware version, where:
yy = last two digits of year
mm = month
dd = day
9.4.2
RS-485 and USB General Read Commands
Command
Response
Description
R1
xx
Read units, where xx is:
PA = Pa
TR = Torr
MB = mbar
R3
xx
Read resolution, where xx is:
LO = low
HI = high
R5
xyzzabb
Read analog output range, where:
x = 1 (linear), 2 (log), 3 (linear-by-decade)
y = high value exponent sign
zz = high value exponent
a = low value exponent sign
bb = low value exponent
R6
xx
Read gas type, where xx is:
N2 = Nitrogen
H2 = Hydrogen
NE = Neon
C2 = Carbon Dioxide
AR = Argon
HE = Helium
KR = Krypton
R7Y
ppseePPSEE
Read SPL (ppsee) and SP2H (PPSEE) settings:
Y = 1 (Open Collector)
Y = 2 (Relay)
R8
x
Read sensor voltage control, where x is:
1 = menu control
2 = pin control
R10
x
Read high voltage state, where x is:
0 = off
1 = on