Instruction Manual
MX4A Active Convection Gauge
3.4
RS-485 Sample Commands
Table 4: RS-485 Command Examples
Command
Output
Description
*0R1<CR>
0001
Units are in Pa
0002
Units in Torr
0003
Units in mbar
*0R2<CR>
ppsePPSE
Lower set point 1 given by ppse
Upper set point 1 given by PPSE
*0R3<CR>
ppsePPSE
Lower set point 2 given by ppse
Upper set point 2 given by PPSE
*0W10001<CR>
0001
Units are in kPa
*0W10002<CR>
0002
Units are in Torr
*0W10003<CR>
0003
Units are in mbar
*0W2ppsePPSE<CR>
ppsePPSE
Assign set point 1 lower to ppse and
Assign set point 1 upper to PPSE
*0W2ppsePPSE<CR>
ppsePPSE
Assign set point 2 lower to ppse and
Assign set point 2 upper to PPSE
*0W4xx<CR>
GG
Change gas type: N2=Nitrogen/Air, AR=Argon
*0S1<CR>
ppse
Vacuum reading is ppse
*0RC1<CR>
Baaa
Read Vac. Adjustment
*0RC2<CR>
Baaa
Read 1 Torr Adjustment
*0RC3<CR>
Baaa
Read 70 Torr Adjustment
*0RC4<CR>
Baaa
Read Atm. Adjustment
*0WC1Baaa<CR>
PPSE
Vac. Adjustment is Baaa, reading is PPSE
*0WC2Baaa<CR>
PPSE
1 Torr Adjustment is Baaa, reading is PPSE
*0WC3Baaa<CR>
PPSE
70 Torr Adjustment is Baaa, reading is PPSE
*0WC4Baaa<CR>
PPSE
Atm. Adjustment is Baaa, reading is PPSE