TELEVAC MX200 Setup And Troubleshooting Manual Download Page 1

 

© 2017 The Fredericks Company | www.frederickscompany.com | [email protected]

 

| +1 215 947 2500

 

televac_an3018 rev - 

Page 

1

 of 

12 

 

 

 

 

 

Televac Application Note 3018 

MX200 EthernetIP Vacuum Controller 

Setup and Troubleshooting Guide for EthernetIP Option 2 

 

 

 

Contents 

Description....................................................................................................................................................... 1 

Windows™ 7 Operating Systems Configuration ................................................................................................ 2 

Windows™ 10 Operating Systems Configuration .............................................................................................. 7 

Contact Us ..................................................................................................................................................... 12 

 

 

 

1

 

Description 

 

 

This document explains the process for the setup, configuration, and troubleshooting of a Televac EthernetIP device. 

In order to communicate with the unit, the EthernetIP device’s IP address will need to be configured to the same 

network as the PLC. The following steps describe how to establish the initial connection, and what steps to follow 

should any issues arise. The second setup option assumes no internet connectivity and explains how to reconfigure 

the IP address of the local computer to match the module, which can then be connected to in order to adjust the 

settings to the appropriate configuration. If an internet connection is available, refer to the Setup and 

Troubleshooting Guide for Televac EthernetIP Option 1 (televac_an3017). 

 

Please see the EthernetIP for MX200 User Manual for a full description on Class 1 and Class 3 connections, and a 

complete ADI table including data instance, name, array length, access, and description. Please see Application 

Note 3016 for downloading an Electronic Data Sheet (EDS) file to a PLC. 

 

 

Summary of Contents for MX200

Page 1: ...ed to the same network as the PLC The following steps describe how to establish the initial connection and what steps to follow should any issues arise The second setup option assumes no internet connectivity and explains how to reconfigure the IP address of the local computer to match the module which can then be connected to in order to adjust the settings to the appropriate configuration If an ...

Page 2: ... Configuring the IP Address 2 1 1 Connect the MX200 EthernetIP port on the back of the unit to your computer 2 1 2 Open the Network and Sharing Center in the Control Panel and verify that an Ethernet connection is being made If there is no connection check the cable and reinsert the Ethernet cable into the adapter on the computer 2 1 3 Select the Ethernet Local Area Connection ...

Page 3: ...ales frederickscompany com 1 215 947 2500 televac_an3018 rev Page 3 of 12 2 1 4 In the Local Area Connection Status panel select Properties 2 1 5 In the Local Area Connection Properties panel select Internet Protocol Version 4 TCP IPv4 and select Properties ...

Page 4: ...ll need to be configured to the module to adjust the settings but will need to be brought back onto the local network in order to communicate properly when finished In this example the local machine has an IPv4 address of 149 174 211 52 with a subnet mask of 255 255 0 0 and a default gateway of 149 174 211 2 2 1 7 Adjust the settings of the computer to an IP address of 192 168 0 9 a subnet mask of...

Page 5: ...ted by the module should appear Select the Configuration tab on the left 2 1 10 Change the IP address and subnet mask to the desired configuration see step 2 1 6 for the network configuration In this example step 2 1 6 showed that the network needed is 149 174 211 X with a subnet mask of 255 255 0 0 The unit in this example will be configured to an IP address of 149 174 211 99 which has been verif...

Page 6: ...igured network connectivity can be verified through the PLC or through the web functionality To verify through the web functionality type the IP address into address bar in the browser and enter the username and password default Admin admin 2 2 2 If the unit is connected to the network properly the web page should be displayed From here data parameters can be viewed and changed along with current ...

Page 7: ...on 3 1 Configuring the IP Address 3 1 1 Connect the MX200 EthernetIP port on the back of the unit to your computer 3 1 2 Open the Network and Sharing Center in the Control Panel and verify that an Ethernet connection is being made If there is no connection check the cable and reinsert the Ethernet cable into the adapter on the computer 3 1 3 Select the Ethernet Connection ...

Page 8: ...company com sales frederickscompany com 1 215 947 2500 televac_an3018 rev Page 8 of 12 3 1 4 In the Ethernet Status panel select Properties 3 1 5 In the Ethernet Properties panel select Internet Protocol Version 4 TCP IPv4 and select Properties ...

Page 9: ...l need to be configured to the module to adjust the settings but will need to be brought back onto the local network in order to communicate properly when finished In this example the local machine has an IPv4 address of 149 174 211 116 with a subnet mask of 255 255 0 0 and a default gateway of 149 174 211 2 3 1 7 Adjust the settings of the computer to an IP address of 192 168 0 9 a subnet mask of...

Page 10: ...ted by the module should appear Select the Configuration tab on the left 3 1 10 Change the IP address and subnet mask to the desired configuration see step 3 1 6 for the network configuration In this example step 3 1 6 showed that the network needed is 149 174 211 X with a subnet mask of 255 255 0 0 The unit in this example will be configured to an IP address of 149 174 211 99 which has been verif...

Page 11: ...igured network connectivity can be verified through the PLC or through the web functionality To verify through the web functionality type the IP address into address bar in the browser and enter the username and password default Admin admin 3 2 2 If the unit is connected to the network properly the web page should be displayed From here data parameters can be viewed and changed along with current ...

Page 12: ...om 1 215 947 2500 televac_an3018 rev Page 12 of 12 4 Contact Us Please feel free to contact us with any questions The Fredericks Company Televac 2400 Philmont Avenue Huntingdon Valley PA 19006 USA Web www frederickscompany com Email sales frederickscompany com Phone 1 215 947 2500 ...