![TELEVAC MC300 Instruction Manual Download Page 21](http://html.mh-extra.com/html/televac/mc300/mc300_instruction-manual_1082535021.webp)
3
CONFIGURATIONS
3
Configurations
The MC300 can be configured to run two low vacuum sensors and an optional high vacuum sensor.
The low vacuum sensors available are the 2A Thermocouple sensor and the 4A Thermocouple Convection
sensor. The two low vacuum sensors must be the same type. The optional high vacuum sensors are
the 7B Penning Magnetron cold cathode, 7E/F/FC Double Inverted Magnetron cold cathode, and the
Mini Bayard-Alpart sensor.
Note that the 7FCS Quick Start Double Inverted Magnetron cold cathode can be operated by both the
2-4502-303 and 2-4502-304 configurations of the MC300 with the Quick Start feature disconnected.
Table 8: MC300 Configuration Part Numbers
Model
Part Number
MC300 2A, 7B
2-4502-301
MC300 4A, 7B
2-4502-302
MC300 2A, 7E/F/FC
2-4502-303
MC300 4A, 7E/F/FC
2-4502-304
MC300 2A, Mini BA
2-4502-305
MC300 4A, Mini BA
2-4502-306
MC300 2A only
2-4503-401
MC300 4A only
2-4503-402
If you are uncertain of your configuration, you can check the code that is displayed upon start up
against Table 9. Note that
Y.Y
refers to the revision number of the software.
Table 9: MC300 Startup Software Code
Code
Instrument Type
Y.Y 2H
2A and Mini BA Sensor
Y.Y 4H
4A and Mini BA Sensor
Y.Y 2b
2A and 7B Sensor
Y.Y 2F
2A and 7E/7F/7FC Sensor
Y.Y 4b
4A and 7B Sensor
Y.Y 4F
4A and 7E/7F/7FC Sensor
3.1
Sensor Descriptions
3.1.1
2A Thermocouple Sensor
The 2A thermocouple sensor has a measurement range from 1
×
10
−
3
up to 20 Torr. It indirectly
measures absolute pressure by detecting the amount of energy that is thermally dissipated from a
heated filament.
3.1.2
4A Thermocouple Convection Sensor
The 4A convection sensor has a measurement range from 1
×
10
−
3
to 1000 Torr. The sensor features
a rapid response time when compared to typical thermal sensors. It indirectly measures absolute
pressure by detecting the amount of energy that is thermally dissipated from a heated filament. It
provides an extended range in comparison to other thermocouple or pirani sensors by utilizing the
physical principle of convection.
email: [email protected]
tel: +1 215 947 2500
web: frederickscompany.com
mc300_im rev H
20 of 23