TelePost LP-100A Operation Manual Download Page 11

 

11 

Detailed Operation 

Cont’d

  

 

Bargraph Decay

. Allows setting the decay time of the power bargraph. The default setting is Fast, and provides a decay of less than 

one second. The longest setting is Slow, and provides about a 3 second decay. In all cases, the attack is instantaneous. 

 

Coupler Type.

 This screen is used to select different maximum power values to be used with custom high power couplers.  Use Dn/Up 

to cycle through the choices. The default is LPC1 (the standard coupler

). Current choices are “LPC1 3KW 1.8-54MHz, “LPC2 5KW 1.8-

30MHz”, “LPC3 250W 0.1 - 20MHz”, ”, “LPC4 5KW 1.8-54MHz”,  “LPC5 10KW 1.8-30MHz” and “LPC6 1KW 0.10 to 10MHz”. Use 
Dn/Up to cycle through choices. 
 

SWR Resting Style.

 

This screen is used to select t

he way you want SWR displayed when you are not transmitting. The choices are… “-

.--

“, “1.00”, “. . . .”, blank and hold last SWR reading. If you select Hold Last, it will be reset when you transmit again. Use Dn/Up to cycle 

through choices. 
 

Lower Bargrph Mode.

 This screen is used to select what parameter is displayed on the lower half of the display. The choices are SWR 

and Reflected Power.  If you select Reflected Power, remember that the reflected power will be referenced to either NET power or 
Forward Power (F+R) depending on your earlier selection for the power display. F+R is the preferred choice to use with REF pwr. Use 
Dn/Up to select. 
 

Callsign Entry

. This screen is used to program your callsign into the screen saver. The Dn button is used to select the position of the 

letter you want to change… 1 thru 6 from left to right. The Up button is used to scroll through the choices… 0 thru 9, A thru Z, space, / 
and -.  Both buttons wrap around. Step thru the positions, scrolling to the letter you want for each position. The callsign is saved as you 
see it. 
 

Display Brightness.

 This screen is used set the display brightness. Each step represents a 12.5% change in brightness. The default 

setting is 6, which equals a brightness level of 75%. This provides almost full brightness, and provides some measure of added display 
life. You can use any brightness level you like. The display is rated for 50,000 hours (5.7 years) of continuous display at full brightness 
before brightness drops to half. With the LP-100A

’s screen savers, you can expect much more than that with typical operating habits. 

Use Dn to reduce brightness, Up to increase. The brightness of the screen changes as you adjust it. 
 

Screen Save Timers.

 This screen is used set the display screen savers. The two timers that can be set are the Scroll timer and the 

Sleep timer. The Scroll timer sets the time in minutes from the last transmission to the time when your call sign starts scrolling across 
the screen. The Sleep timer sets the time from the last transmission to the time when the display turns off. The Scroll saver should be 
set first, since it also affects the Sleep timeout. Each can be adjusted for up to 10 minutes (20 minutes total). The screen saver extends 
display life, and reduces power consumption and heat when the meter is idle. There is also a third screen saver timer, but it is factory 
preset. It dims the screen to 25% one second after transmission ends when in the Main Mode. If Peak power mode is selected and the 
hold time is set for 1 second or more, it dims at the end of the hold period.  
 
 

 
Calibration 
 

Calibration Initial Screen.

 This screen simply identifies that you are in the Calibration mode. 

 

Serial Number.

 All LP-100As after serial #100 are supplied with RG-58U connecting cables between the coupler and main chassis. 

Earlier versions used RG-174U. This screen allows selection of the appropriate cable. It selects the proper correction table for the cable 
loss vs. frequency. On later versions, this screen name was changed to Serial Number to compensate for other hardware changes as 
well as cable type. 
 

Gain Zero Trim.

 This screen allows band-by-band calibration of the balance of the gain detector. The process simply requires a good 

quality dummy load. The Dn/Up buttons are adjusted until the resistance on the screen matches the resistance of your dummy load. 
The LP-100A automatically saves the Cal constants for each band, indexed to frequency. The built-in frequency counter automatically 
determines the frequency. 
 

Phase Zero Trim.

 This screen allows band-by-band calibration of the balance of the phase detector. The Dn/Up buttons are adjusted 

until the phase on the screen reads zero degrees. The LP-100A automatically saves the Cal constants for each band, indexed to 
frequency. The built-in frequency counter automatically determines the frequency. 
 

Gain Slope Trim.

 Allows setting the slope of the magnitude for proper Z at a value removed from 50 ohms. This can be done with any 

reasonable known load in the 25 or 75-100 ohm range. I am also working on a calibrator kit to simplify this. 
 

Phase Slope Trim.

 Allows calibrating the phase detector. This requires a delay line of known value. In its simplest form, this can be 

done by calculating the electrical length of an existing piece of coax in the 3-10

’ range, and matching the readout to the calculated 

length at the frequency used for the calculation. More on this in the Calibration section. I am also working on a calibrator kit to simplify 
this. 
 

Summary of Contents for LP-100A

Page 1: ...1 LP 100A Digital Vector RF Wattmeter Operations Manual Covers LP 100A serial numbers starting at 2020 and firmware beginning at v1 2 3 1...

Page 2: ...igan Radiation Laboratory Industry Canada Compliance Statement Canada Digital Apparatus EMI Standard This Class B digital apparatus meets all the requirements of the Canadian Interference Causing Equi...

Page 3: ...verview 6 Detailed Operation 9 Setup Menus 14 Circuit Description 16 Schematic 18 Software 20 Specifications 26 Warranty 27 Appendix 28 Calibration 31 Note Calibration info is provided for reference L...

Page 4: ...d in the field to external standard to within 0 1 on each band Power display is Fwd or Net power delivered to the load Fwd minus Ref power SWR Alarm system with set points for Off 1 5 2 0 2 5 3 0 and...

Page 5: ...T amp keying between your amplifier and rig through the LP 100A using RCA connectors RS 232 Connects to computer standard M F straight through DB9 serial cable See manual for usage Current Voltage Con...

Page 6: ...n as well as using the Plot program to do automatic graphing of a number of parameters dBm mode uses professional dBm and RL Return Loss terminology instead of watts and SWR to indicate power and load...

Page 7: ...e Peak Mode is VERY fast and can respond to a lip smack mic button click etc Don t be alarmed by this it is normal and allows the LP 100A to provide an accurate indication of peak power Unless a lot o...

Page 8: ...es This is especially true of older amplifiers with unregulated power supplies but also is affected by the ALC timing characteristics of modern rigs in both CW and SSB The peak detector in the LP 100A...

Page 9: ...except that it is calibrated to display power from 15 dBm to 33 dBm There is no return loss in this mode because it does not utilize the coupler Power is supplied directly to one of the inputs on the...

Page 10: ...in this mode resets the PIC quite useful when flash programming the PIC The reset does not affect setup calibration settings Pressing the Peak Avg Tune button toggles the Temp display between degrees...

Page 11: ...imer and the Sleep timer The Scroll timer sets the time in minutes from the last transmission to the time when your call sign starts scrolling across the screen The Sleep timer sets the time from the...

Page 12: ...nd SSB The peak detector in the LP 100A is very fast and will grab even the smallest peak Peak SWR will show values a little higher than steady state at times due to the wide dynamic range of the LP 1...

Page 13: ...There is no return loss in this mode because it does not utilize the coupler Power is supplied directly to either one of the inputs on the back of the LP 100A This mode can be used for accurate low p...

Page 14: ...10 above the indicated value Note Defaults are 15W 100W and 1500W Note These ranges are scaled by a factor of x1 67 when using a 5KW coupler and 3 33 when using a 10KW coupler This screen is used to s...

Page 15: ...me from the last transmission to the time when the display turns off The Scroll saver should be set first since it also affects the Sleep timeout Each can be adjusted for up to 10 minutes 20 minutes t...

Page 16: ...which provide signals to both the gain phase comparator and the power detector With a 50 ohm non reactive load the levels of these two signals will be virtually identical and the phase between them wi...

Page 17: ...itches interfacing to the GVFD display processor and an SWR alarm relay which is used to kill the PTT to your amplifier to protect both the antenna and amplifier The SWR alarm also lights a front pane...

Page 18: ...18 Schematic Page 1...

Page 19: ...19 Schematic Page 2 Coupler Schematic...

Page 20: ...to download a different driver than the one supplied with this unit when using XP Search the internet for ICUSB232 ByteRunner also sells an adapter which uses the well supported FTDI chipset for about...

Page 21: ...15 200 baud 8 bits no parity 1 stop bit NOTE Firmware versions before 1 2 0 0 used a baud rate of 38 400 and before 1 0 3 used a baud rate of 19 200 and did not report dBm or SWR values MicroCode Load...

Page 22: ...d Ham Radio Deluxe HRD Transmit mode is also selectable For most rigs FSK is a good choice but AM and CW are also available To use CW you will probably need an interface which uses the RTS or DTR hand...

Page 23: ...tart frequency and subtracts 1 kHz form the stop frequency This is necessary because some rigs will not transmit at the band edges in all modes Set a step size For single band scans of most bands 50 k...

Page 24: ...How can you tell the sign is bad Generally speaking X or Phase never bounce off of zero or swing radically through zero like from 20 to 20 over a 50 kHz span If this happens reverse the first point af...

Page 25: ...of the feedline to the remote antenna would be needed or the LP 100A located remotely with a laptop computer which would allow remote control of the LP 100A To use PolarPlot with the LP 100A you just...

Page 26: ...Bargraph response 50 Hz Bargraph resolution 90 steps Direct Inputs 15 to 33 dBm 50 ohms 0 1 to 650 MHz 0 1 5 dB 2W max DC Power 11 16 VDC 270 330 mA depends on brightness setting Operating temp range...

Page 27: ...ation of the warranty period at our discretion or where in our reasonable opinion the damage is due to abuse accident improper or abnormal usage improper installation alteration lightning or other inc...

Page 28: ...such as the HP438A that we use which has a calibrated precision thermocouple sensor These sensors are generally designed to operate in the 0 1 to 1W range This necessitates the need for a precision p...

Page 29: ...especially at the leading edge of a CW character or voice syllable This behavior also depends on the resting time between characters The effect is exaggerated when the SWR is high since the rig will t...

Page 30: ...ubs sometime in the future For a comprehensive review of the principles and math involved in transmission line theory line transformations etc visit http www cebik com trans zcalc html This is copyrig...

Page 31: ...d and setting the Trim for a reading of 25 0 or whatever the actual load resistance is if it s not exactly 25 0 This adjustment is used to calibrate the slope of the phase detector It is simply done b...

Page 32: ...hes your dummy load then unkey the transmitter Repeat this procedure for all bands Advance to the Phase Zero screen and repeat the procedure for each band this time adjusting for a phase of zero degre...

Page 33: ...s for the adjustment of Master power sensitivity This value will normally be near 1 Both the Master and Fine power adjustments have a range of 12 5 in 0 1 increments The Fine sensitivity adjustment is...

Page 34: ...and the variation of Fine Trim setting should be 2 from 160 10m and 10 on 6m This is dependent on a number of factors to do with xfmr winding mainly total of core wound positioning and wire routing an...

Reviews: