7.1.17.
TGC (Time Gain Compensation)
The Time Gain Compensation (TGC) controls adjust how returning signals are amplified in order to
correct the attenuation that varies depending on the scanning depth and the tissue being observed. Using
the TGC controls you can equalize amplification of the returning echo signal so that the density of echoes at
different depths is similar. TGC allows you to control the gain level in several depth zones independently.
TGC control points are spaced proportionally to the current scanning depth. Small circular markers on this
curve correspond to the TGC control points.
Values
TGC values available are in the interval 0-100% (in software they are not shown). Measurement
units used - percentages (%).
Adjustment
When scanning controls are not active, press and hold for several seconds on ultrasound image (at
any depth) at the right of depth scale line to show TGC curve. Then, while TGC curve is visible, do horizontal
sweeps on ultrasound image at desired to change TGC control points (depth) that are painted as small
circular markers. Software will automatically hide TGC curve when it is not changed for some time.
Relations to other controls
To adjust the overall gain use the B mode Gain control. After adjusting the Depth you may need to
adjust the TGC.