background image

 

 

TFO-480A 

© 2010 –

 

2012 Telefire Fire & Gas Detectors Ltd 

 

Revision 1.01   January 2012 

 

– Page 4 of 5 – 

   Note    

Please check the detector’s value in the control panel’s MONITOR screen.  If 
the value is higher than 103, replace the detector.  The detector should be 
cleaned if it is lower than 61. 

4.1 

Periodic Testing 

Fire alarm systems should be checked periodically.  Please refer to NFPA and local fire 
codes to determine service frequency. 
Use the control panel’s Walk Test mode to perform automatic reset.  Please refer to the 
control panel’s manual for additional details on how to perform Walk Test. 

4.1.1  Test Procedure – Locally 

1.  Apply a magnet to the side of the detector next to the LED for 3 – 5 seconds.  This 

activates an internal testing circuit that simulates presence of smoke in the detector, 
tests the sensing mechanism and the detector's electronic circuit. 

   Warning    

Do not apply naked flame to the detector! 

2.  The analog value representing the smoke level will be transmitted to the control 

panel for evaluation.  The control panel will transmit a signal to the detector to turn 
on its LED.  During the test the detector’s values may be observed at the control 
panel.  Please refer to the ADR-3000 manual for additional information. 

3.  Once the detector is in alarm mode, it keeps the alarm condition until reset by the 

control panel.  If the control panel is in "Walk Test" mode, it will reset the alarm after 
a few seconds.  Please refer to the ADR-3000 manual for instructions on how to 
conduct a walk test. 

4.1.2  Test Procedure – From the Control Panel  

It is also possible to test the detector using the ADR-3000’s “Monitor” option menu.  
Please refer to the ADR-3000 manual for instructions on how to use and interpret the 
monitor screen. 

   Note    

Testing detectors with an aerosol spray may cause increased contamination 
in the labyrinth chamber due to accumulation of oil residue and dust.  This 
accumulation may shorten the detectors life-span and require it to be sent to 
the factory for cleaning. 

4.2 

Detector Cleaning 

Contamination is a by-product of normal operation and may be caused by various 
sources, most of which are impossible to eliminate.  Normal human activity creates a 
constant source of airborne dust and dirt that over a period of time may affect the 
detector's performance. 
Analog detectors should be cleaned when a maintenance signal appears on the control 
panel.  Local regulations may specify an interval period for maintenance schedule. 
The TFO-480A's chamber is removable and can be cleaned or replaced.  Cleaning 
should be done by qualified personnel in order to ensure that the detector is properly 
handled, reassembled correctly and tested for proper operation. 

Summary of Contents for TFO-480A

Page 1: ...nical Manual TELEFIRE FIRE GAS DETECTORS LTD PO Box 7036 Petach Tikva 49250 Israel Tel 972 3 921 1955 Fax 972 3 921 1816 eMail marketing telefire co il Web www telefire co il TFO TFO TFO TFO 480A 480A...

Page 2: ...line UL 864 and UL 268 standards and Fault as used in EN 54 standards are used interchangeably throughout this manual i Note Do not install operate and maintain this product before fully reading this...

Page 3: ...smoke from burning of various materials Soft set address the detector s address is programmed into its memory without the use of mechanical switches or moving parts Can be tested directly with a test...

Page 4: ...ressable Detector and Accessory Programmer Please refer to the PROG 4000 manual for additional details 3 2 2 ADR 3000 Configuration Configure the detector as Photoelectric Detector in the ADR 3000 Con...

Page 5: ...see the control panel s technical manual for an explanation on performing Detectors Init i Note Ensure that detector initialization is performed when all detectors are in a smoke free condition 3 3 Po...

Page 6: ...detector is in alarm mode it keeps the alarm condition until reset by the control panel If the control panel is in Walk Test mode it will reset the alarm after a few seconds Please refer to the ADR 3...

Page 7: ...3 non condensing Sensitivity Range set at control panel 0 8 2 0 foot obscuration Operating Voltage supplied by control panel via SLC 21V Modulated Maximum Current Consumption Quiescence mode 290 A Ala...

Reviews: