Read the Operators Manual thoroughly before using the instrument.
;
Calibrate once per shift, and prior to each and every use.
;
Check the high and low alarm threshold settings prior to use.
;
Replace the batteries immediately when required.
;
Keep the instrument, cable and sensor dry at all times.
;
Re-calibrate after changing the sensor or batteries.
;
Visually inspect the whole instrument prior to use.
;
Check the integrity of all cable connections routinely.
;
Remove and carefully store the “flow diverter” when using the instrument in “static”
environments such as incubators, oxygen hoods etc.
;
Perform an alarm functionality test prior to each use.
;
Remove batteries if the instrument is not to be used for a month plus
D
ON
’
TS
:
Use the Accurox Oxygen Monitor if you suspect any malfunction.
:
Use the monitor in the presence of flammable anaesthetics.
:
Autoclave the instrument and/or the sensor.
:
Submerge the instrument and/or sensor in water or any other liquid.
:
Expose instrument to other equipment with high levels of RFI emission.
:
Place the instrument itself in a water vapour saturated environment.
:
Expose the LCD to excessive direct sunlight
:
Dispose of the sensor cell in fire, or place in an incinerator
.