3.7.2. Warm Up
During the warm-up period various portions of the instrument’s front panel may
behave as indicated in Table 3-11.
Table 3-11: Front Panel Display During System Warm-Up
NAME
COLOR
BEHAVIOR
SIGNIFICANCE
Conc field
N/A
Displays current,
compensated CO
2
Concentration
This is normal operation, but deemed inaccurate during the
warm-up period.
Mode field
N/A
Displays current
mode, e.g.,
SAMPLE”
Instrument is in Sample mode but is still in the process of
warming up. (DAS holdoff period is active)
Param field
N/A
Displays menu items
and messages.
Refer to Table 3-12 for a list and descriptions of warning
messages.
STATUS LEDs
Sample Green
On
Unit is operating in sample mode; front panel display is
being updated.
Flashes On/Off when adaptive filter is active
Cal Yellow
Off
The
instrument’s calibration is not enabled.
Fault Red
Blinking
The analyzer is warming up and hence out of specification
for a fault-free reading. Various warning messages appear
in the Param field.
3.7.3. Warning Messages
Because internal temperatures and other conditions may be outside be specified
limits during the analyzer’s warm-up period, the software will suppress most
warning conditions for 30 minutes after power up. If warning messages persist
after the 30 minutes warm up period is over, investigate their cause using the
troubleshooting guidelines in Section 8 of this manual.
To view and clear warning messages, press:
SAMPLE
HVPS WARNING CO2 = 0.00
TEST
CAL
MSG
CLR
SETUP
Make sure warning messages are
not due to real problems.
Press
CLR
to clear the current
message.
If more than one warning is active, the
next message will take its place
Once the last warning has been
cleared, the analyzer returns to
SAMPLE
mode
SAMPLE
RANGE=500.000 PPM
CO2 = 0.00
< TST TST > CAL
MSG
CLR
SETUP
SAMPLE
HVPS WARNING
CO2 = 0.00
TEST CAL
MSG
CLR
SETUP
TEST
deactivates warning
messages
MSG
activates warning
messages.
<TST TST
> keys replaced with
TEST
key
NOTE:
If the warning message persists
after several attempts to clear it,
the message may indicate a
real problem and not an artifact
of the warm-up period
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...