Operating Instructions
Teledyne API Model T360/T360M Operation Manual
136
The following DC current output limits apply to the current loop modules:
Table 4-17: Analog Output Current Loop Range
RANGE
MINIMUM OUTPUT
MAXIMUM OUTPUT
0-20 mA
0 mA
20 mA
These are the physical limits of the current loop modules, typical
applications use 2-20 or 4-20 mA for the lower and upper limits. Please
specify desired range when ordering this option.
The default offset for all ranges is 0 mA.
Pin assignments for the output connector at the rear panel of the instrument are
shown in Table 4-18.
ANALOG OUT
A1 A2 A3 A4
+ - + - + - + -
Table 4-18: Analog Output Pin Assignments
PIN
ANALOG
OUTPUT
VOLTAGE
SIGNAL
CURRENT
SIGNAL
1
V Out
I Out +
2
A1
Ground
I Out -
3
V Out
I Out +
4
A2
Ground
I Out -
5
V Out
I Out +
6
A3
1
Ground
I Out -
7
V Out
Not used
8
A4
Ground Not
used
1
Output A3 is only used when the
O
2
sensor option is installed
See Figure 3-4 for the location of the analog output connector on the instrument’s
rear panel.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...