Teledyne API – Model T300/T300M CO Analyzer
Theory of Operation
319
13.4.5.3.
THERMISTOR INTERFACE
This circuit provides excitation, termination and signal selection for several negative-
coefficient, thermistor temperature sensors located inside the analyzer. They are as
follows:
S
AMPLE
T
EMPERATURE
S
ENSOR
The source of this signal is a thermistor located inside the sample chamber of the Optical
Bench. It measures the temperature of the sample gas in the chamber. This data is used
to during the calculation of the CO concentration value.
B
ENCH
T
EMPERATURE
S
ENSOR
This thermistor is attached to the sample chamber housing. It reports the current
temperature of the chamber housing to the CPU as part of the bench heater control loop.
W
HEEL
T
EMPERATURE
S
ENSOR
This thermistor is attached to the heatsink on the GFC Wheel motor assembly. It reports
the current temperature of the wheel/motor assembly to the CPU as part of the Wheel
Heater control loop.
B
OX
T
EMPERATURE
S
ENSOR
A thermistor is attached to the motherboard. It measures the analyzer’s internal
temperature. This information is stored by the CPU and can be viewed by the user for
troubleshooting purposes via the front panel display (see Section 12.1.2).
13.4.5.4.
ANALOG OUTPUTS
The analyzer comes equipped with four analog outputs:
A1
,
A2
,
A3
and
A4
. The type
of data and electronic performance of these outputs are configurable by the user (see
Section 5.4).
O
UTPUT
L
OOP
-
BACK
All four analog outputs are connected back to the A/D converter through a loop-back
circuit. This permits the voltage outputs to be calibrated by the CPU without need for
any additional tools or fixtures.
13.4.5.5.
INTERNAL DIGITAL I/O
This channel is used to communicate digital status and control signals about the
operation of key components of the Optical Bench. The CPU sends signals to the
sync/demod board that initiate the
ELECTRICAL TEST
and
DARK CALIBRATION
procedures.
13.4.5.6.
EXTERNAL DIGITAL I/O
This External Digital I/O performs two functions: status outputs and control inputs.
S
TATUS
O
UTPUTS
Logic-Level voltages are output through an optically isolated 8-pin connector located on
the rear panel of the analyzer. These outputs convey good/bad and on/off information
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...