Teledyne API – Model T300/T300M CO Analyzer
Theory of Operation
317
13.4.4.5.
STATUS LEDS
Eight LEDs are located on the analyzer’s relay board to show the current status on the
various control functions performed by the relay board. They are listed on Table 13-4.
Table 13-4: Relay Board Status LEDs
LED
COLOR
FUNCTION
STATUS WHEN LIT
STATUS WHEN UNLIT
D1
RED
Watch Dog Circuit
Cycles On/Off every 3 seconds under direct control of the analyzer’s
CPU.
D2
YELLOW
Wheel Heater
HEATING
NOT HEATING
D3
YELLOW
Bench Heater
HEATING
NOT HEATING
D4
YELLOW
Spare
N/A
N/A
D5
GREEN
Sample/Cal Gas Valve
Option
Valve Open to CAL GAS FLOW
Valve Open to SAMPLE Gas Flow
D6
GREEN
Zero/Span Gas Valve Option
Valve Open to SPAN GAS FLOW
Valve Open to ZERO GAS FLOW
D7
GREEN
Shutoff Valve Option
Valve Open to CAL GAS FLOW
Valve CLOSED to CAL GAS
FLOW
D8
GREEN
IR SOURCE
Source ON
Source OFF
Figure 13-15: Location of relay board Status LEDs
13.4.4.6.
I2C WATCH DOG CIRCUITRY
Special circuitry on the relay board monitors the activity on the I
2
C bus and drives LED
D1. Should this LED ever stay ON or OFF for 30 seconds, the watch dog circuit will
automatically shut off all valves as well as turn off the IR Source and all heaters. The
GFC Wheel motor will still be running as will the Sample Pump, which is not controlled
by the relay board.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...