Setup Menu
Teledyne API – Model T300/T300M CO Analyzer
124
5.9.3.2.
ANALOG OUTPUT CALIBRATION
Analog output calibration should to be carried out on first startup of the analyzer
(performed in the factory as part of the configuration process) or whenever recalibration
is required. The analog outputs can be calibrated automatically, either as a group or
individually, or adjusted manually.
In its default mode, the instrument is configured for automatic calibration of all
channels, which is useful for clearing any analog calibration warnings associated with
channels that will not be used or connected to any input or recording device, e.g.,
datalogger.
Manual calibration should be used for the 0.1V range or in cases where the outputs must
be closely matched to the characteristics of the recording device. Manual calibration
requires the AUTOCAL feature to be disabled.
Automatic calibration can be performed via the
CAL
button located inside The
AOUTS
CALIBRATION
submenu. By default, the analyzer is configured so that calibration of
analog outputs can be initiated as a group with the
AOUT CALIBRATION
command.
The outputs can also be calibrated individually, but this requires the AUTOCAL feature
be disabled.
5.9.3.3.
ENABLING OR DISABLING THE AUTOCAL FOR AN INDIVIDUAL ANALOG OUTPUT
To enable or disable the
AutoCal
feature for an individual analog output, elect the
ANALOG I/O CONFIGURATION
submenu (see Figure 5-3) then press:
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...