Getting Started
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
34
3.3.1.2. CONNECTING ANALOG INPUTS (OPTION)
The Analog In connector is used for connecting external voltage signals from other
instrumentation (such as meteorological instruments) and for logging these signals in the
analyzer‟s internal DAS. The input voltage range for each analog input is 0-10 VDC and
input impedance is nominally 20kΩ in parallel with 0.1µF.
Figure 3-7:
Analog In Connector
Pin assignments for the Analog In connector are presented in Table 3-4.
Table 3-4:
Analog Input Pin Assignments
PIN
DESCRIPTION
DAS
PARAMETER
1
1
Analog input # 1
AIN 1
2
Analog input # 2
AIN 2
3
Analog input # 3
AIN 3
4
Analog input # 4
AIN 4
5
Analog input # 5
AIN 5
6
Analog input # 6
AIN 6
7
Analog input # 7
AIN 7
8
Analog input # 8
AIN 8
GND
Analog input Ground
N/A
1
See Section 7 for details on setting up the DAS.
3.3.1.3. CONNECTING ANALOG OUTPUTS
The rear panel Analog Output channels
A1
,
A2
and
A3
are assigned to the NO
x
, NO and
NO
2
concentration signals of the analyzer with a default output voltage setting of 0 to 5
VDC and a reporting range of 0 to 500 ppb.
A4
is assigned a user-selected diagnostic test function (see Section 5.9.4), also with a 0
to 5 VDC default analog output voltage setting.An optional Current Loop output is
available for
A1
,
A2
and
A3
only.
To access these signals attach a strip chart recorder and/or data-logger to the appropriate
analog output connections on the rear panel of the analyzer. Pin-outs for the analog
output connector are:
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...