3.3. CONNECTIONS AND SETUP
This section presents the electrical (Section 3.3.1) and pneumatic (Section 3.3.2)
connections for setup and preparing the instrument for operation.
3.3.1. ELECTRICAL CONNECTIONS
Note
To maintain compliance with EMC standards, cable length must be no greater
than 3 meters for all I/O connections, which include Analog In, Analog Out, Status
Out, Control In, Ethernet/LAN, USB, RS-232, and RS-485.
3.3.1.1. CONNECTING POWER
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying
at least 10 A current at your AC voltage and that it is equipped with a functioning earth
ground.
WARNING
E
LECTRICAL
S
HOCK
H
AZARD
High Voltages are present inside the analyzers case.
Power connection must have functioning ground connection.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting or connecting electrical
subassemblies.
Do not operate with cover off.
CAUTION
G
ENERAL
S
AFETY
H
AZARD
To avoid damage to your analyzer, ensure that the AC power voltage
matches the voltage indicated on the analyzer’s model/specs label (Figure
3-4) before plugging the T200 into line power.
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...