Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Calibration Procedures
197
9.4.3.3. USE OF ZERO/SPAN VALVE WITH REMOTE CONTACT CLOSURE
Contact closures for controlling calibration and calibration checks are located on the rear
panel
CONTROL IN
connector. Instructions for setup and use of these contacts are
found in Section 3.3.1.6.
When the contacts are closed for at least 5 seconds, the instrument switches into zero,
low span or high span mode and the internal zero/span valves will be automatically
switched to the appropriate configuration.
The remote calibration contact closures may be activated in any order.
It is recommended that contact closures remain closed for at least 10 minutes to
establish a reliable reading.
The instrument will stay in the selected mode for as long as the contacts remain
closed.
If contact closures are being used in conjunction with the analyzer‟s AutoCal (see
Section 9.5) feature and the AutoCal attribute “
CALIBRATE”
is enabled, the T200 will
not re-calibrate the analyzer UNTIL when the contact is opened. At this point, the new
calibration values will be recorded before the instrument returns to
SAMPLE
mode.
If the AutoCal attribute “
CALIBRATE”
is disabled, the instrument will return to
SAMPLE
mode, leaving the instrument‟s internal calibration variables unchanged.
9.5. AUTOMATIC ZERO/SPAN CAL/CHECK (AUTOCAL)
The AutoCal system allows unattended periodic operation of the Z
ERO/SPAN
valve
options by using the T200‟s internal time of day clock. AutoCal operates by executing
SEQUENCES
programmed by the user to initiate the various calibration modes of the
analyzer and open and close valves appropriately. It is possible to program and run up
to three separate sequences (
SEQ1
,
SEQ2
and
SEQ3
). Each sequence can operate in
one of three modes, or be disabled.
Table 9-2:
AUTOCAL Modes
MODE NAME
ACTION
DISABLED
Disables the Sequence.
ZERO
Causes the Sequence to perform a Zero calibration/check.
ZERO-SPAN
Causes the Sequence to perform a Zero point
calibration/check followed by a Span point
calibration/check.
SPAN
Causes the Sequence to perform a Span concentration
calibration/check only.
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...