9.3.2. SETUP FOR CALIBRATION WITH THE INTERNAL SPAN GAS
GENERATOR
Connect the sources of zero air and span gas as shown in Figure 9-4.
VENT
here if input
is pressurized
Source of
SAMPLE GAS
(reconnected after
“precision” cal
performed)
PUMP
Enclo
sure Wall
External Zero
Air Scrubbe
r
Filter
Chassis
SAMPLE
EXHAUST
ZERO AIR
FROM DRYER
Figure 9-4:
Pneumatic Connections for Manual Calibration/Checks with the Internal Span Gas
Generator
9.3.3. CAL ON NO
2
FEATURE
When using the IZS option to calibrate the T200, the analyzer‟s
CAL_ON_NO
2
feature
must be turned on. This feature enables a continuous zero gas flow across the IZS
permeation tube and through the NO
2
converter. It also programs the analyzer to use the
NO output from the NO
2
converter to calibrate the span value of both NO and NO
X
.
Note
This feature should only be enabled when a span calibration or calibration check
is performed.
While
CAL_ON_NO
2
is enabled, the NO
2
concentration will always be reported as
zero. This is because the gas is continuously routed through the NO
2
converter and the
analyzer‟s firmware simulates calibration with NO gas.
Table 9-1:
IZS Option Valve States with CAL_ON_NO
2
Turned ON
Valve
Condition
Valve Port Connections
Sample/Cal
Open to zero/span valve
1
2
Zero/Span
Open to SPAN GAS inlet
1
2
NO/NO
x
Valve
Open to NO
2
converter
1
2
Auto Zero Valve
Cycles normally
N/A
Since the instrument sees the same concentration of NO during both NO and NO
X
cycles, it reports an NO
2
concentration of zero.
06858E DCN7057
Summary of Contents for T200
Page 2: ......
Page 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Page 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Page 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Page 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Page 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Page 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Page 393: ...06858E DCN7057 D 7 ...