Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Getting Started
47
The status outputs are accessed via a 12-pin connector on the analyzer’s rear panel
labeled STATUS (Figure 3-9). Pin-outs for this connector are presented in Table 3-7.
STATUS
1 2 3 4 5 6 7 8 D
+
S
Y
S
T
E
M
OK
OK
H
IG
H
R
A
N
G
E
C
O
N
C
V
A
L
ID
Z
E
R
O
C
A
L
S
P
A
N
C
A
L
D
IA
G
M
O
D
E
Figure 3-9: Status Output Connector
Table 3-7: Status Output Signals
REAR PANEL
LABEL
STATUS
DEFINITION
CONDITION
1
SYSTEM OK
ON if no faults are present.
2
CONC VALID
OFF any time the
HOLD OFF
feature is active, such as during calibration or when
other faults exist possibly invalidating the current concentration measurement
(example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid.
3
HIGH RANGE
ON if unit is in high range of either the DUAL or Auto range modes.
4
ZERO CAL
ON whenever the instrument’s
ZERO
point is being calibrated.
5
SPAN CAL
ON whenever the instrument’s
SPAN
point is being calibrated.
6
DIAG MODE
ON whenever the instrument is in
DIAGNOSTIC
mode
7&8
SPARE
D
EMITTER BUS
The emitters of the transistors on pins 1-8 are bussed together.
SPARE
+
DC POWER
+ 5 VDC, 300 mA source (combined rating with Control Output, if used).
Digital Ground
The ground level from the analyzer’s internal DC power supplies
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...