Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
274
06807A(DCNxxxx)
J2: Power
Configuration
Jumper
Figure 10-4:
Location of Relay Board Power Configuration Jumper
AC Configuration of the pump is accomplished via an in-line, hard wired, set of
connections. Call Teledyne ML’s Customer Service Department for more information.
10.6.2. DC POWER SUPPLY
If you have determined that the analyzer’s AC main power is working, but the unit is still
not operating properly, there may be a problem with one of the instrument’s switching
power supplies, which convert AC power to 5 and ±15 V (PS1) as well as +12 V DC
power (PS2). The supplies can either have DC output at all or a noisy output (fluctuating).
To assist tracing DC Power Supply problems, the wiring used to connect the various
printed circuit assemblies and DC powered components and the associated test points on
the relay board follow a standard color-coding scheme as defined in Table 10-4.
Table 10-4: DC Power Test Point and Wiring Color Code
NAME
TEST POINT#
COLOR
DEFINITION
DGND
1
Black
Digital ground
+5V
2
Red
AGND
3
Green
Analog ground
+15V
4
Blue
-15V
5
Yellow
+12V
6
Purple
+12R
7
Orange
12 V return (ground) line
A voltmeter should be used to verify that the DC voltages are correct as listed in Table
10-4. An oscilloscope, in AC mode and with band limiting turned on, can be used to
evaluate if the supplies are excessively noisy (>100 mV peak-to-peak).
Summary of Contents for T100
Page 2: ......
Page 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Page 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Page 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Page 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Page 21: ...21 PART I GENERAL INFORMATION ...
Page 77: ...77 PART II OPERATING INSTRUCTIONS ...
Page 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Page 207: ...207 PART III TECHNICAL INFORMATION ...
Page 208: ...208 06807A DCNxxxx ...
Page 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Page 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Page 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...