PS200
PORTABLE GAS MONITOR
RECHARGING THE BATTERIES
64171, Rev. 11/December 2019
7-1
7. RECHARGING THE BATTERIES
7.1. General
Use only approved chargers to recharge monitor.
The battery should be recharged when:
• ‘LOW BATTERY’ or ‘BAT FAULT’ appears on the display.
• The monitor will not switch on.
7.2. Recharge Monitor Using the Charging / Comms Clip
1. Connect the supplied Charging / Comms Clip to the monitor. Ensure the locating tongue on the
clip engages with the monitor locating slot and is firmly seated (refer to
Charging/Comms Clip Connection
).
Figure 7-1: Charging / Comms Clip Connection
2. Connect the supplied MINI-USB to USB Cable to the Charging / Comms Clip.
WARNING: CHARGING IS ONLY PERMITTED IN A NON-HAZARDOUS AREA.
CAUTION: Switch the monitor off when charging.
Contact Pads (x4)
Charging / Comms Clip (P/N 64260)
Locating Slot
Locating Tongue