36
64171LMP
Revision 3
PS200-LMP
PORTABLE GAS MONITOR
Appendix A. Typical Operating Parameters
Gas
Sensor Type
Range
Resolution
Accuracy
% VOL Gas
(Natural Gas)
Catalytic/Thermal
Conductivity
0 to 100%
0.1%
±3%
Carbon Monoxide (CO)
Electrochemical
0 to
1000ppm
1ppm
±10%
(of indication)
Physical Properties
Dimensions
(H x W x D):
4.8” x 2.3” x 1.3” (121mm x 59mm x 32mm)
Weight:
8 oz. (230 g)
Environment
Operating
Temperature Limits:
-4ºF to +122ºF (-20ºC to +65ºC)
Storage Temperature
Limits:
-40ºF to +149ºF (-40ºC to +65ºC)
Humidity:
0 to 95% R.H non-condensing
Typical Flow Rate
Information:
Nominal pump flow rate is 0.132 to 0.185 gals./min. (0.5 to 0.7 liters/min.)
Max. 97 ft. (30m) sample line.
Typ. flow fail rate = 0.026 to 0.053 gals./min. (0.1 to 0.2 liters/min.)
Response Times
Vol Gas
T50: <10s (CSA 22.2 No. 152)
T60: <12s (ANSI 12.13.01)
CO
T90: <20s
Warm-up/Stabilization Time
<40 seconds
Alarms
Highly visible, flashing LED; piercing >90dB audible
Display
Green/Red LCD backlit display
Power Source
Battery:
Rechargeable Lithium-Ion
Runtime:
Up to 24 hours (12 hours pumped)
Charging Time:
Up to 4 hours
Construction
High-impact, rubberized polycarbonate case
IP Rating
IP67