Chapter 8
Troubleshooting
125
8.2
Acquisition Started but Viewer Contents Do Not Update
In LightField, live data is normally displayed in the Experiment workspace viewer as it is
being acquired (Preview or Acquire mode). If the viewer is not being updated and
acquisition is occurring, check to see if there is a filename in the top row of the viewer:
in
, the filename
2010 May 11 13_13_49.raw
is displayed. If there is a
filename, click on the camera icon. The file data will be cleared from the viewer and the
live data will then be displayed.
Figure 8-2: Acquisition Display
8.3
Baseline Signal Suddenly Changes
A change in the baseline signal is normal if the temperature, gain, or speed setting has
been changed. If this occurs when none of these settings have been changed, there
may be excessive humidity in the camera vacuum enclosure. Turn off the camera and
contact Teledyne Princeton Instruments Customer Support. Refer to
on page 160 for complete information.
4
4
11-
0126
_0
080
Summary of Contents for ProEM Series
Page 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Page 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Page 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 161: ...This page is intentionally blank ...