Chapter 7
Troubleshooting
87
7.13 Secure the Detector-Controller Cable Slide Latch
Some Teledyne Princeton Instruments Detector-Controller cables use a slide latch to
secure the Detector-Controller cable to the DETECTOR connector on the back of the
ST-133. Incorrectly plugging this cable into the connector and improperly securing the
slide latch may prevent communication with the PI-MTE (the camera may appear to
stop working).
Perform the following procedure to connect and secure the Detector-Controller cable:
1. Before attempting to plug in the cable, slide the latch up toward Pin 1 as illustrated
in
Figure 7-13: Connector View of Unlatched Detector Connector with Pin 1
2. Plug the cable into the DETECTOR connector on the ST-133.
Figure 7-14: Typical Unlatched Detector Connector
4
4
11-
009
7_
004
8
P
IN
1
U
NLATCHED
4
4
11
-009
7_
004
9
F
ULLY
S
LIDE
L
ATCH
U
P
(UNLATCHED P
OSITION
)
Summary of Contents for Princeton Instruments PI-MTE
Page 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Page 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Page 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Page 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Page 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 117: ...This page is intentionally blank...