Chapter 4
System Setup
37
5. The Welcome dialog similar to that illustrated in
is displayed.
Figure 4-3: Camera Detection Wizard - Welcome Dialog
Verify I will configure the camera system manually is unchecked. If necessary,
deselect it.
6. Click Next.
7. Follow on-screen prompts to perform the initial hardware setup.
This wizard configures default parameters on the Hardware Setup dialog and
provides the opportunity to acquire a test image to confirm the system is working.
NOTE:
Note that this is only a test image and it is not acquired
using the settings needed for true data acquisition.
For a step-by-step procedure about basic system operation,
refer to
4.11 Make the Coolant Connections
CAUTION!
!
1. Take care that the coolant used is pH neutral. Acidic or
alkaline coolant can damage the camera fittings and
internal cooling block through corrosion. Such damage
could be very expensive to repair.
2. Coolant should be no colder than +10° C to +15°C to
prevent condensation at 50% relative humidity. Operating
the camera with coolant at a colder temperature could
cause induced condensation in the electronics enclosure
and possible catastrophic damage to the camera when the
camera is operating outside of a vacuum enclosure.
Damage resulting from this type of operation may void the
warranty.
4
4
11-
009
7_
001
7
Summary of Contents for Princeton Instruments PI-MTE
Page 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Page 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Page 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Page 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Page 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Page 117: ...This page is intentionally blank...