Appendix E:
Troubleshooting
WARNING!
Do not attach or remove any cables while the camera
system is powered on.
E.1
General Camera Faults/Errors
This section provides information about troubleshooting general camera faults and
errors.
E.1.1
Connection Failure or Logic Power Supply Overcurrent
This failure is indicated by the red FAULT LED on the rear of the Lansis power supply
being ON (i.e., does not flash.)
Examine the primary power cable and repair/replace as necessary.
Verify all connectors are fully seated at both ends and the jack screws are secured.
E.1.2
Power Supply Switch in On Position, But Power LED Extinguished
If the power supply has been turned on (i.e., power switch is in the I position,) but the
green POWER LED does not illuminate indicating the power supply is operational:
•
Turn the power switch off (i.e., the 0 position);
•
Wait approximately five seconds;
•
Turn the power switch on (i.e., the I position)
Table E-1: Troubleshooting Index by Error/Fault Description
Error/Fault
Refer to…
Power Supply Switch in On Position, But Power LED Extinguished
Temperature Lock Cannot be Achieved or Maintained
Gradual Deterioration of Cooling Capability
External Coolant Circulator: Low Coolant (Air in the Hoses)