76
Lansis
®
System Manual
Issue 1
EAR99
Technology
Subject
to
Restrictions
Contained
on
the
Cover
Page.
A.2.1
Thermal Characteristics
Refer to
for specific thermal information.
Imaging Area
30.72 mm x 3.96 mm
Spectra/s
b
(full vertical
binning)
~236 fps
ADC Speed/16 bits
100 kHz, 2 MHz
Vertical Shift Rate
c
7.41
sec/row
Read Noise
5 e
–
rms @ 100 kHz (typical)
Optical Focal Distance
TBD
to the Lansis data sheets available for download from
www.princetoninstruments.com
.
b. @ 2 MHz full vertical binning
c. Software programmable
Table A-3: Lansis CCD Specifications
a
(Sheet 2 of 2)
Specification
A-261BR
A-261BRX
W-261BR
W-261BRX
Table A-4: Default Operating Temperature
Specification
Cooling Medium
Specification
Default Operating Temperature
Air
-60°C
Liquid
a
a. External chiller required.
-65 °C
Deepest Cooling Temperature
Air
-65 °C
Liquid
-70 °C
Precision
–
±0.05 °C