CONNECTIONS AND STARTUP
This section presents the electrical (Section 2.3.1) and pneumatic (Section 2.3.2)
connections for setting up and preparing the instrument for operation (Section 2.3.3).
ELECTRICAL CONNECTIONS
Note
To maintain compliance with EMC standards, cable length
must be no greater than 3 meters for all I/O connections.
WARNING – Electrical Shock Hazard
•
High Voltages are present inside the instrument’s case.
•
Power connection must have functioning ground connection.
•
Do not defeat the ground wire on power plug.
•
Turn off instrument power before disconnecting or connecting
electrical subassemblies.
•
Do not operate with cover off.
CAUTION – Avoid Damage to the Instrument
Ensure that the AC power voltage matches the voltage indicated on
the instrument’s rear panel before plugging it into line power.
C
ONNECTING
P
OWER
ATTENTION
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Never power off the instrument from the rear panel Hard
Power switch before first powering down the internal
components through the front panel Soft Power switch.
Attach the power cord between the instrument’s AC power connector and a power outlet
capable of carrying at least the rated current at your AC voltage range and ensure that it is
equipped with a functioning earth ground. It is important to adhere to all safety and
cautionary messages.