CombiFlash® R
f
Installation Guide
2-2
WARNING
The system is heavy. Use a two-person lift to prevent injury.
Before making any connections to the Combi
Flash
R
f
, place
the system on the bench or in the fume hood where it will be
operated. Temporarily position the system so you can access
the back panel to complete the connections.
2.3 Connect Power
Ensure that the On/Standby switch below the touch screen
panel is in the Standby position. Then, use the supplied IEC
power cord to connect the Combi
Flash
R
f
to mains power.
WARNING
Mains power must meet the voltage, frequency, and amperage
requirements listed on the serial number label.
2.4 Connect Solvent Lines
WARNING
Risk of fire ignited by electrostatic discharges. Never
substitute the black tubing on CombiFlash systems. The black
PTFE tubing (P/N 023-0503-06) is conductive. This tubing is
required to dissipate static electricity.
CAUTION
To prevent damage or premature wear to the pump and internal
valves, clean solvent should be used. The solvent should not
contain any dissolved solids.
The Combi
Flash
R
f
has four solvent inlet ports on the back
panel: two A solvents and two B solvents. The system requires
at least one A and one B solvent to form a binary gradient, and
assumes that the B solvent is stronger than A. The second set
of inputs allow you to form alternative binary gradients, or
change the A or B solvent mid-run.
Accessory package (P/N 60-5239-006) includes several pieces
to complete the solvent inlet connections.