HDO9000 High Definition Oscilloscopes Operator's Manual
Setting Up WaveScan
This procedure explains how to set up WaveScan to search an acquisition for events of interest. Set
up your source channel and triggers before setting up the scan.
1. Press the front panel
Stop button
to stop acquisition.
2. Choose
Analysis > WaveScan
.
3. Check
Enable
.
4. Choose the
Source
waveform.
5. Choose the
and enter values for any additional settings that appear at the right of
the dialog based on your selection.
6. If you're using Measurement Mode, set up the filter in one of the following ways:
l
Touch
Filter
and choose an operator, then enter the
Filter Limit
.
l
Touch
Filter Wizard
and choose one of the pre-set filters. The Filter and Filter Limit are
automatically set based on your selection.
7. Select each
in which you wish to display results by checking the box at the bottom
of the dialog. Each view selected is displayed simultaneously.
8. If you're using Scan Overlay view, on the Scan Overlay dialog
Clear Sweeps
. If desired, set up
the
display.
9. Optionally, choose an
Action
to trigger when an event that meets your scan criteria is found.
10. Restart acquisition.
11. When using the Zoom view, use the Z
x
tab to adjust the zoom.
Scan Overlay
To apply monochromatic persistence to a Scan Overlay:
1. Select
Scan Overlay
when setting up the wave scan, then open the
Scan Overlay dialog
.
2. Check
Persistence On
.
3. Enter a
Saturation
level as a percentage. All samples above the saturation level are assigned
the highest color intensity.
4. Choose a
Persistence Time
. The higher the time, the more static the persistence display.
To adjust the scan overlay to "zoom" in or out: touch the
In/Out buttons
, or touch
Scale
and enter new
values. Check
Var.
to adjust scale in finer steps than the default 1, 2, 5.
128