PORTABLE GAS MONITOR
30
64171
6.5.2 Bump Test – Fail
If any range fails the bump test, the CONFIRM ALARMS display will be red and a
symbol will
be displayed, as shown in
Figure 47: Confirm Alarms - Fail
.
Figure 47: Confirm Alarms - Fail
After selecting
YES
or
NO
, BUMP TEST FAIL is displayed and the user is instructed to switch the
monitor off, as shown in
Figure 48: Bump Test Fail - Switch Monitor OFF
.
Figure 48: Bump Test Fail - Switch Monitor OFF
If a monitor fails a bump test, then a full re-calibration is necessary.
Revision 12
.1