MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE 

 (Addendum to GFC7000E Manual PN 04584)

05134 Rev -

6

Warning

Message

Fault Condition

Possible Causes

FRONT PANEL

WARN

The CPU is unable to

Communicate with

the Front Panel

Display /Keyboard

WARNING

 only appears on Serial I/O COM Port(s)

Front Panel Display will be frozen, blank or will not
respond.

 

Failed Keyboard

 

I

2

C Bus failure

 

Loose Connector/Wiring

PHOTO TEMP

WARNING

PHT DRIVE

 is

>4750 mVDC or

<250 MVDC

 

Failed IR Photo-Detector

 

Failed Sync/Demod Board

 

IR Photo-Detector improperly attached to the

Sample Chamber

 

BENCH TEMP

 too high.

REAR BOARD

NOT DET

Mother Board not

detected on power

up.

WARNING

 only appears on Serial I/O COM Port(s)

Front Panel Display will be frozen, blank or will not
respond.

 

Massive Failure of Mother Board

RELAY BOARD

WARN

The CPU cannot

communicate with

the Relay Board.

 

I

2

C Bus failure

 

Failed Relay Board

 

Loose connectors/wiring

PURGE PRESS

WARN

Purge Pressure

outside of warning

limits specified by

PURGE_PRESS_SET

variable

 

Loss of purge flow

 

Failed purge pressure sensor

 

Purge pressure regulator not set properly

SAMPLE PRES

WARN

Sample Pressure is

<15 in-Hg or

 > 35 in-Hg

Normally 29.92 in-

Hg at sea level

decreasing at 1 in-

Hg per 1000 ft of

altitude (with no

flow – pump

disconnected).

If Sample Pressure is < 15 in-HG:

 

Blocked Particulate Filter

 

Blocked Sample Inlet/Gas Line

 

Failed Pressure Sensor/circuitry

If Sample Pressure is > 35 in-HG:

 

Blocked Vent line on pressurized

Sample/Zero/Span gas supply

 

Bad Pressure Sensor/circuitry

 

Failed Permapure dryer

SAMPLE TEMP

WARN

Sample temperature

is < 10

o

C or >

100

o

C.

 

Ambient Temperature outside of specified

range

 

Failed Bench Heater

 

Failed Bench Temperature Sensor

 

Relay controlling the Bench Heater

 

Failed Relay Board

 

I

2

C Bus

SOURCE

WARNING

Occurs when CO Ref

is <1250 mVDC or

>4950 mVDC.

Either of these

conditions will result

in an invalid M/R

ratio.

 

GFC Wheel stopped

 

Failed Sync/Demod Board

If Status LED’s on the Sync/Demod Board 

ARE

flashing the cause is most likely a failed:

 

IR Source

 

Relay Board

 

I

2

C Bus

 

IR Photo-Detector

SYNC

WARNING

Phase Lock Loop

(PLL) has lost lock
on wheel rotation.

 

GFC Wheel stopped

 

GFC Wheel rotation very slow/dragging

Summary of Contents for GFC7000EU

Page 1: ...584 05134 Rev 1 MODEL GFC7000EU C O2 ANALYZER Addendum to GFC7000E Manual PN 04584 Teledyne Analytical Instruments 16830 Chestnut St City of Industry Ca 91748 TEL 626 961 9221 FAX 626 961 2538 TEL 626 934 1500 E MAIL ai customerservice teledyne com WEB SITE http www teledyne ai com ...

Page 2: ...rence manual covering the details of the instrument s components Only the sections of the manual that are different between the GFC7000E and a GFC7000EU are listed in this addendum 2 SPECIFICATIONS AGENCY APPROVALS AND WARRANTY 2 1 Specifications Table Model GFC7000EU Specifications Ranges User selectable to any full scale range from 0 100 ppb to 0 100 ppm Measurement Units ppb ppm µg m3 mg m3 use...

Page 3: ...us Outputs 8 Status outputs from opto isolators Control Inputs 6 Control Inputs 2 defined 4 spare I O One 1 RS 232 One 1 RS 485 RS 232 Ethernet Baud Rate 300 115200 Certifications CE EN61010 1 90 A1 92 A2 95 EN61326 Class A 1 As defined by the USEPA 2 At constant temperature and voltage 3 Or 0 2 ppm whichever is greater 4 Or 0 1 ppm whichever is greater 5 Above 10 ppm range otherwise 0 2 ppm for l...

Page 4: ...centration limits are exceeded 11 1 1 Interpreting WARNING Messages The most common and or serious instrument failures will result in a warning message being displayed on the front panel The table below lists warning messages along with their meaning and recommended corrective action It should be noted that if more than two or three warning messages occur at the same time it is often an indication...

Page 5: ...zer s CPU or Mother Board can result in any or ALL of the following messages Table Warning Messages Warning Message Fault Condition Possible Causes BENCH TEMP WARNING The optical bench temp is controlled at 48 5 C Bench Heater Bench Temperature Sensor Relay controlling the Bench Heater Entire Relay Board I2 C Bus BOX TEMP WARNING Box Temp is 5 C or 48 C Box Temperature typically runs 7o C warmer t...

Page 6: ...RESS_SET variable Loss of purge flow Failed purge pressure sensor Purge pressure regulator not set properly SAMPLE PRES WARN Sample Pressure is 15 in Hg or 35 in Hg Normally 29 92 in Hg at sea level decreasing at 1 in Hg per 1000 ft of altitude with no flow pump disconnected If Sample Pressure is 15 in HG Blocked Particulate Filter Blocked Sample Inlet Gas Line Failed Pressure Sensor circuitry If ...

Page 7: ...er Fatal Error caused software to restart Loose connector wiring WHEEL TEMP WARNING The filter wheel temperature is controlled at 68 5 C Blocked Cooling Vents below GFC Assembly Analyzer s Top Cover removed Wheel Heater Wheel Temperature Sensor Relay controlling the Wheel Heater Entire Relay Board I2 C Bus APPENDIX B GFC7000EU Spare Parts List NOTE Use of replacement parts other than those supplie...

Page 8: ...M300E B17 033560000 Field Mirror 32 Pass M300E 035950100 Disk On Chip with Software 036110400 Sync Demodulator Board 036310000 4 20 mA Output PCA 037250000 Assembly Heater Optical Bench 039750000 Keyboard 040010000 ASSY Fan Rear Panel E Series 040030300 PCA Press Sensors w SW49 035280000 CO2 Scrubber Assembly 040690200 Mother Board relay alarm outputs 040880000 Wheel Position Sensor Assembly 04135...

Page 9: ...V 039060000 Assy Pressure Transducer M360U GFC7000 Table B 2 GFC7000E Expendables List Part Number Description Qty 009600400 M300E Expandables Kit Includes This kit contains the following items 009690100 Akit TFE Fltr Elements 47mm 5um 25 1 FL0000001 Sintered Filter 002 024900 1 HW0000020 Spring Flow Control 1 OR0000001 O Ring Flow Control 2 PU0000022 Pump Rebuild Kit KNF Model NO5ATI 1 Note01 23 ...

Reviews: