B-6
FIELD CALIBRATION
FIRST RESPONDER CG - USER HANDBOOK
4. Turn the regulator valve counter clockwise to
open the valve slightly. Make sure that the gas is
flowing before connecting the sample tubing to the
instrument, otherwise an instrument sample fault
may occur.
5. Connect tubing from regulator to instrument inlet
then adjust the regulator valve to maintain a constant
flow of gas (counter clockwise to increase flow and
clockwise to decrease). The correct flow rate is
achieved when the ball in the indicator tube floats
just above its resting position.
6. Wait for the instrument gas reading to settle.
7. If the displayed reading corresponds to the
concentration of calibration gas, i.e. 50% LEL (2.5%
Methane in Air), proceed to paragraph 10.
8. If the displayed reading does not correspond to the
concentration of calibration gas, i.e. 50% LEL (2.5%
Methane in Air), press and hold
(Button 1) to
enter SPAN mode.
SPAN mode is indicated by the selected range, in
this case LEL, and SPN alternating in the display as
shown in Figure B-6.
Fig. B-6 SPAN Mode Display
Summary of Contents for First Responder CG
Page 1: ...First Responder CG User Handbook ...
Page 14: ...1 4 FIRST RESPONDER CG USER HANDBOOK ...
Page 48: ...3 22 FIRST RESPONDER CG USER HANDBOOK ...
Page 66: ...5 14 FIRST RESPONDER CG USER HANDBOOK ...
Page 72: ...8 2 FIRST RESPONDER CG USER HANDBOOK ...
Page 76: ...A 4 FIRST RESPONDER CG USER HANDBOOK ...
Page 88: ...C 4 FIRST RESPONDER CG USER HANDBOOK Fig C 3 Semiconductor Sensor Flexi Probe Applications ...
Page 96: ...C 12 FIRST RESPONDER CG USER HANDBOOK ...
Page 102: ...VI FIRST RESPONDER CG USER HANDBOOK ...
Page 103: ......