2101 Field Wizard
Section 4 Modbus Protocol
4-8
(1) A write to the Identify module register will cause the module to perform the identify operation which may be a steady
LED for a few seconds or a beep in the Field Wizard.
(2) Setting the Take Reading flag to 1 will cause the module to update the registers with current data readings. It will
be set to zero when the readings have all been updated. This may be used to initiate readings and poll for when
they are ready to be read. It may take up to 50 seconds to update all the readings, depending upon the flow con-
ditions. Setting the Take Reading flag to 2 causes an automatic, 15 second update of readings when a Modbus
master is polling the 2100.
(3) The Update Interval specifies an interval in seconds that the registers are automatically updated. It defaults to zero,
which indicates that no automatic updating will occur.
(4) The Active Flag (1-4) bit fields specify what fields/registers are active in the list. This provides support for a maxi-
mum of 64 fields. For example, if bit 0 of register 27 is set, the Level (registers 40,41) is active. If bit 1 of register
27 is set, then the Velocity (registers 55,56) is active. If bit 0 of register 28 is set, the Analog channel 7 (registers
265,266) is active.
(5)A non-zero status code indicates a measurement problem.
(6) Time is represented in a series of registers: Order is from lowest address to highest - Seconds (0-59), Minutes (0-59),
Hours (0-23), Days (1-31), Month (1-12) and Year (1977-2099).
117
Volume 1 status code
16-bit integer
Read
118-123
Volume 1 time record
Time
Read
130,131
Voltage
4-byte float
Volts
Read
132
Voltage status code
16-bit integer
Read
133-138
Voltage time record
Time
Read
145,146
Temperature
4-byte float
Degrees Celsius
Read
147
Temperature status code
16-bit integer
Read
148-153
Temperature time record
Time
Read
160,161
Internal Temp
4-byte float
Degrees Celsius
Read
162
Internal Temp status code
16-bit integer
Read
163-168
Internal Temp time record
Time
Read
175,176
Analog channel 1
4-byte float
0-100 percent
Read
177
Analog channel 1 status code
16-bit integer
Read
178-183
Analog channel 1 time record
Time
Read
190,191
Analog channel 2
4-byte float
0-100 percent
Read
192
Analog channel 2 status code
16-bit integer
Read
193-198
Analog channel 2 time Record
Time
Read
205,206
Analog channel 3
4-byte float
0-100 percent
Read
207
Analog channel 3 status code
16-bit integer
Read
208-213
Analog channel 3 time record
Time
Read
220,221
Analog channel 4
4-byte float
0-100 percent
Read
222
Analog channel 4 status code
16-bit integer
Read
223-228
Analog channel 4 time record
Time
Read
235,236
Analog channel 5
4-byte float
0-100 percent
Read
237
Analog channel 5 status code
16-bit integer
Read
Table 4-2 Modbus ASCII Address 2-(N+1) Register Definitions (Continued)
Register Number(s)
Name
Data Type
Units
Read/Write
Summary of Contents for Field Wizard 2101
Page 2: ......
Page 4: ......
Page 8: ...2101 Field Wizard Table of Contents vi ...
Page 42: ...2101 Field Wizard Section 3 Operation 3 18 ...
Page 52: ...2101 Field Wizard Section 4 Modbus Protocol 4 10 ...
Page 56: ...2101 Field Wizard Section 5 Maintenance 5 4 ...
Page 58: ...2101 Field Wizard Appendix A Replacement Parts A 2 ...
Page 59: ...2101 Field Wizard Appendix A Replacement Parts A 3 ...
Page 60: ...2101 Field Wizard Appendix A Replacement Parts A 4 ...
Page 62: ...2101 Field Wizard Appendix B Accessories B 2 ...
Page 70: ......
Page 72: ......
Page 74: ...Warranty ...
Page 76: ......