T3SP-D4MX Operator’s Manual
6.2 Safety Compliance
6.2.1 EC DECLARATION OF CONFORMITY – LOW VOLTAGE
The instrument meets intent of EC Directive 2014/35/EU for Product Safety. Compliance
was demonstrated to the following specifications as listed in the Official Journal of the Eu-
ropean Communities:
EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control,
and laboratory use – Part 1: General requirements
EN 61010-2:030:2010 Safety requirements for electrical equipment for measurement, con-
trol, and laboratory use – Part 2-030: Particular requirements for testing and measuring
circuits
The design of the instrument has been verified to conform to the following limits put forth
by these standards:
•
Mains Supply Connector: Overvoltage Category II, instrument intended to be supplied
from the building wiring at utilization points (socket outlets and similar).
•
Measuring Circuit Terminals: No rated measurement category. Terminals not intended
to be connected directly to the mains supply.
•
Unit: Pollution Degree 2, operating environment where normally only dry, non-conduc-
tive pollution occurs. Temporary conductivity caused by condensation should be ex-
pected.