Information on Charging the Battery
CEA9001 is equipped with an NiCad storage battery. The service life and capacity of the battery are considerably
affected by the way the instrument is charged and used. In order to make the handling safer, the instrument has a
load management unit.
If an NiCad battery is, for example, always charged from 80% to 100% and never run down to the final discharge
voltage, it will lose some of its capacity. This is called the 'memory effect' , i.e. the battery remembers to what
extent it is run down.
A part of this memory effect is suppressed in the
in that the battery cannot be recharged until it has
dropped below 60%.
Constant overcharging, too, has adverse effects on the NiCad battery. In order to prevent this, the charged
capacity, the voltage and the temperature of the battery are monitored in the
. When predefined limits
are exceeded, the charging process is interrupted. After the appropriate parameters have been neutralized the
charging process is automatically restarted again.
The service life of the NiCad battery can be significantly reduced when the instrument is operated at
temperatures below 40 °F (5°C).
The graphic charge-level indicator of the
(10 battery symbols), which appears in the one-line status
display during the calibration phase, helps the user estimate correctly the capacity of the battery. The instrument
continuously measures the incoming and outgoing current during operation and charging. Under normal
operating conditions, the instrument should be operated until the battery is completely run down. When this
advice is followed, the actual capacity of the NiCad battery will definitely be shown on the display.
CEA9001
CEA9001
CEA9001
Summary of Contents for CEA 9001
Page 1: ... US Version V3 1 C E A ombustion fficiency nalyzer US Version V3 1 ...
Page 3: ......
Page 13: ...10 4 Operating Elements 0 I START STOP S1 S2 S3 ENTER mbar hPa ENTER ...
Page 14: ...11 ...
Page 17: ...14 Natural Gas ...
Page 18: ...15 ...
Page 23: ...20 Cursor flashes OK to key in your entry ...
Page 24: ...Intermediate storage of the graphics and allmeasured values 21 ...
Page 26: ...23 ABC abc0123 ...
Page 28: ...25 No Files are existend ...
Page 30: ...27 ...
Page 31: ...28 ...
Page 35: ...32 ...
Page 36: ...33 Attention All memories inlcuding additional data will be deleted ...
Page 38: ...35 Adjustable display illumination from 0 90 seconds in increments of 5 seconds ...
Page 45: ...42 ...
Page 46: ...43 ...