2–11
Operational Theory 2.0
Teledyne Analytical Instruments
A Business Unit of Teledyne Electronic Technologies
Power for the control module is provided by a center-tapped trans-
former which takes the 115 VAC input, reduces it to 40 VAC, and feeds the
voltage to a DC power supply identical to the one installed in the detector
module. Power supply outputs are +24 VDC unregulated, and +15 VDC
regulated.
Ext.
Volt.
Amplifier
Alarm
Comparator
Auto
Zero
Timer
Manual
Setpoint
Adjust.
#1
#2
Relays
N.O.
Com.
N.C.
N.O.
Com.
N.C.
E-to-I
Converter
Current
(4
to
20
mA)
Standard
Valve
Control
Signal
Span
Adjust.
Meter
Indication
Voltage
(0
to
1
V
f.s.)
Millivolts
(0
to
5
mV
f.s.)
To
I-to-E
Converter
To Timer
15 VDC
24 VDC
Power
Supply
Power
Transformer
Manual
Fine
Zero
Power
115
VAC
60
Hz
Input
I-to-E
Converter
Current
Signal
Input
From
Analysis
Unit
115
VAC
60
Hz
Output
±
±
Figure 2-5.
Control Module - Block Diagram