
06160J DCN7969
Teledyne API Model 465M O
3
Monitor User Manual
53
Figure 9-2: Instrument Layout, NEMA Configuration
9.3.
Replacing Internal Particulate Filter
1.
Disconnect power from the 465M.
2.
Loosen the six screws from the top cover (Rack Mount Configuration, see Figure
9-1) or open front panel (NEMA Configuration, see Figure 9-2.)
3.
Remove the six screws and top cover from the instrument.
4.
Locate the particulate housing filter on the rear panel (Rack Mount Configuration,
see Figure 9-1) or the bottom panel (NEMA Configuration, see Figure 9-2.)
5.
Loosen the four screws on the sample filter body (see Figure 9-3.)
DRYER