![Teledyne 401 Instruction Manual Download Page 53](http://html.mh-extra.com/html/teledyne/401/401_instruction-manual_1080808053.webp)
Teledyne API Model 401 O
3
Photometric Calibrator Instruction Manual, 01124, Rev. J1
Table 4-1: Setup Variables
Button Sequence
Function
Default
Limits
SETUP-EVNT-WAIT
Zero cal. duration
Span cal. duration
15 min.
15 min.
1-20 min.
1-20 min.
SETUP-EVNT-TIME Cal.
time-of-day
23:30
00:00-23:59
SETUP-EVNT-ASEQ
Auto. Multipoint cal. enable
OFF
OFF/ZS/Z2S/Z5S
SETUP-EVNT-SHFT
Auto check shift
0 min.
±
60 min.
SETUP-COMM-BAUD
RS-232 baud rate
2400 baud
300, 1200, 2400
SETUP-COMM-ID
Analyzer ID number
0000
0000-9999
SETUP-MISC-O3-SLPE
Slope adjust parameter
1.00
0.85 - 1.15
SETUP-MISC-O3-GEN-ADJ O
3
GEN lamp setup/adjust
OFF
ON - OFF
SETUP-MISC-O3-GEN-CAL O
3
GEN calibration
OFF
ON - OFF
SETUP-MISC-O3-GEN-
MODE
Feedback mode selection
BNCH
SETUP-MISC-O3-OFFS
Offset adjust parameter
0.0
±
1000.0
SETUP-MISC-DARK-VIEW
Dark current offset
0 mV
1000 mV
SETUP-MISC-CLK-TIME
Current time-of-day
00:00
00:00 - 23:59
SETUP-MISC-CLK-DATE
Current date
01 JAN 00
31 DEC 99
SETUP-MISC-CLK-ADJ
Clock speed adjust
0 sec/day
±
60 sec.
SETUP-MISC-D/A-RNGE
D/A output range
500 ppb
100-10000
SETUP-MISC-D/A-OFFS
D/A offset setting
0 mV
±
500 mV
SETUP-MISC-MORE-FLT
Fault LED timeout
5 min
0 - 10 min
SETUP-MISC-MORE-UNITS O
3
conc. units
ppb
ppb/ppm/ug/mg
SETUP-MISC-MORE-TCHN
TEST to analog output
OFF
OFF-ON
SETUP-PASS
Cal. password enable
ON
OFF-ON
SETUP-CFG-LIST Software
configuration
O
3
CALIB.
Display
4-7