![Teledyne 3010TAC Operating Instructions Manual Download Page 58](http://html1.mh-extra.com/html/teledyne/3010tac/3010tac_operating-instructions-manual_1080415058.webp)
5 Maintenance
Model 3010TAC Oxygen Analyzer
5-2: Part I
Teledyne Analytical Instruments
3. Replace fuse as shown in Figure 5-2.
4. Reassemble Housing as shown in Figure 5-1.
American Fuses
European Fuses
Figure 5-2: Installing Fuses
5.2
System Self Diagnostic Test
1. Press the
System
button to enter the system mode.
2. Use the
< >
arrow keys to move to
More
, and press
Enter
.
3. Use the
< >
arrow keys to move to
Self-Test
, and press
Enter
.
The following failure codes apply:
Table 5-1: Self Test Failure Codes
Power
0
OK
1
5 V Failure
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed
Summary of Contents for 3010TAC
Page 12: ...iv Part I Model 3010TAC Oxygen Analyzer TeledyneAnalyticalInstruments...
Page 18: ...1 6 Part I 1 Introduction Model 3010TAC Teledyne Analytical Instruments...
Page 22: ...2 4 Part I 2 Operational Theory Model 3010TAC Teledyne Analytical Instruments...
Page 34: ...3 Installation Model 3010TAC 3 12 Part I Teledyne Analytical Instruments...
Page 70: ...2 6 Part II 2 Operational Theory Model 3010TAC Teledyne Analytical Instruments...
Page 76: ...3 Installation Model 3010TAC 3 6 Part II Teledyne Analytical Instruments...