![Teledyne 3010MB Operating Instructions Manual Download Page 36](http://html1.mh-extra.com/html/teledyne/3010mb/3010mb_operating-instructions-manual_1080411036.webp)
4 Operation
Model 3010MB
4-6: Part I
Teledyne Analytical Instruments
TRAK/HLD AutoCal
PSWD Logout More
Use < > arrows to blink
AutoCal
, and press
Enter
. A new screen for
Span/Zero
set appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Press < > arrows to blink
Span
(or
Zero
), then press
Enter
again. (You
won’t be able to set
OFF
to
ON
if a zero interval is entered.) A
Span
Every
... (or
Zero Every
...) screen appears.
Span Every 0 d
Start 0 h from now
Use
∆∆∆∆∆∇
arrows to set an interval value, then use < > arrows to move to
the start-time value. Use
∆∆∆∆∆∇
arrows to set a start-time value.
To turn
ON
the
Span
and/or
Zero
cycles (to activate
Auto-Cal
): Press
System
again, choose
AutoCal
,
and press
Enter
again. When the
Span/
Zero
values screen appears, use the < > arrows to blink the
Span
(or
Zero
)
OFF
/
ON
field. Use
∆∆∆∆∆∇
arrows to set the
OFF
/
ON
field to
ON
. You can
now turn these fields
ON
because there is a nonzero span interval defined.
4.3.3 Password Protection
If a password is assigned, then setting the following system parameters
can be done only after the password is entered: span and zero settings,
alarm setpoints, analysis range definitions, switching between autoranging
and manual override, setting up an auto-cal, and assigning a new password.
However, the instrument can still be used for analysis or for initiating a self-
test without entering the password.
If you have decided not to employ password security, use the default
password
TETAI
. This password will be displayed automatically by the
microprocessor. The operator just presses the Enter key to be allowed total
access to the instrument’s features.
NOTE: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
4.3.3.1
Entering the Password
To install a new password or change a previously installed password,
you must key in and
ENTER
the old password first. If the default password